New W&H sterilizer calibration device

The new W&H sterilizer calibration device has been specially developed for the calibration of the internal pressure and temperature sensors of all W&H sterilizers: Lisa MB, Lisa 300/500, Lisa 500 fully automatic, Lina MB and OEM sterilizers.

When is calibration required?

Calibration of your sterilizer is required whenever the clock chipset has been replaced (LISA MB only) or if required by an Official Authority.

Moreover, calibration is recommended whenever a pressure or temperature sensor, or a printed circuit board has been replaced or if alarms advising out-of range measurement start to go off randomly during the normal use of the sterilizer.


The main features at a glance

  • All in one. All W&H sterilizers can be calibrated with one single device.
  • User-friendly easy handling.
  • Excellent price performance ratio.
  • Built-in memory card for calibration cycles registration.
  • Calibration certificate.

The W&H sterilizer calibration device warranty is one year.