Lara sterilizer

Incredible inside & outside

As unique as your practice
The standard versions of the Lara sterilizers already offer the fastest sterilization cycle in their class – including standard documentation. An additional, newly developed upgrade system offers even more speed and extended documentation if you so wish – and ensures you are optimally equipped even if your requirements change.

Sterilization, Hygiene & Maintenance

Sterilization, Hygiene & Maintenance

System Overview

Lisa



  • Lara incredible outside
    Lara incredible inside

    Lara Sterilizers
    Ready for the future

    Incredible outside
    You don’t need to be a lover of design to appreciate the Lara sterilizers. Their smooth surfaces are comfortable to the touch, the instruments are easy to load and the flat colour screen displays allow you to navigate through the intelligent menu structure quickly and with ease.

    Incredible inside
    Discover the W&H Lara sterilizers down to the smallest detail. So many components, perfectly arranged to offer you complete satisfaction by simplifying the sterilization process while providing you with comprehensive documentation. Unique upgradeability allows you to customise your Lara sterilizer with your personal needs and to comply with regulatory requirements.

    Incredible because

    • Fastest B cycle in its segment
    • Unique upgradeability thanks to Activation Codes
    • Colour touch display for easy navigation
    • Automatic cycle record via high capacity USB stick
    • Optional high end label and cycle report printer
    • Ergonomics and functional design
    • Ideal dimensions to integrate and operate
    • New environmentally friendly packaging
      The new environmentally friendly packaging is completely reusable and recyclable. Plastic components were reduced to a minimum. For 99% of its weight, the new package is made of cardboard and other fully biodegradable substances.
    • Reliable traceability

      Discover simple operation & reliable traceability


      The Lara colour touch screen has been selected and designed to make your daily work easier, faster and more efficient. Lara also offers a comprehensive traceability system customized to your needs.

    • Easy to use

      Easy to use

      The intelligent menu structure of the Lara sterilizers ensures extraordinarily high operating comfort: you can browse to the preset programmes quickly and intuitively using the large colour touchscreen.

    • Data logger

      High capacity data logger

      An high capacity USB drive automatically records the cycle reports throughout Lara’s entire service life. Optionally available: the Lara label and cycle report printer with which you can document the traceability without an additional computer or separate software.

    • User identification

      User identification

      You can also acquire additional features such as user identification with Activation Codes: the user who has loaded/unloaded the sterilizer can be digitally recorded. Users are identified by a 6-digit PIN.


      Discover unique upgradeability

      Whatever occurs in the future, with the Lara sterilizers you’re well prepared.

      Effectively, their innovative upgradability system offers you a cost-efficient, rapid and customized opportunity to activate additional features to reflect your practice needs or comply with future requirements.


      The four additional features at a glance:

      All in One

      With the "All in One" Activation Code you have all benefits in one go – and ultimately your sterilizer comes at a previously unattainable priceperformance ratio.

      Six steps to your customized sterilizer:

      1. Purchase Activation Code of your choice from your local approved dental supplier.
      2. After receiving the related login code go to activation.wh.com
      3. You specify this code along with the serial number and model of your sterilizer.
      4. Your individual, secure Activation Code is generated.
      5. Enter this code in the menu of your Lara sterilizer.
      6. The feature is available.

      W&H Steri App

      W&H Steri App is an advanced smartphone and tablet application that allows to control and monitor the status of the autoclaves in your dental clinic. Moreover, the app is able to scan and track dental instruments within a single sterilizing cycle, keeping a related history database on your smart device.

      Download on the App Store

      Download on Google Play Store

      Compatibility:

      The application is compatible with all sterilizers currently available.
      Available languages: EN, IT, FR, DE, ES, ZH-CN, JP

      Main features:

      Remote control of your sterilizer: the user can remotely command up to 4 sterilizers.
      Advanced traceability: EliTrace feature is now fully supported and the App permits to create instrument database directly from your smart device. In addition to this function, the management and tracking of dental instruments is integrated.
      Reports creation: the App allows making intervention reports in order to collect data regarding specific treatments.
      User management: creation and handling of sterilizer users directly from your smart device.
      Backup functionality: all the cycle reports are automatically saved to assure more protection of data.


      Discover outstanding performance

      The Lara sterilizers offer you one of the fastest type B in its performance class right from the standard version. And you can improve them even by activating the additional features Eco Dry + and Fast Cycle.

      Eco Dry  + automatically adapts the drying time to the load – and, in doing so, saves time and energy. Additionally available via an Activation Code: a type S fast cycle for unpackaged, semicritical products.

      Reprocessing in less than 23 minutes!


      Lara sterilizer

      supplied with reversible rack, 3 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test

      Lara 17
      REF 19941120

      Chamber size 17 l

      Lara 22
      REF 19942120

      Chamber size 22 l

      Buy

      Download Area

      FAQs

      • What is validation?
        Validation is a testing procedure carried out by an official authority, proving the effectiveness of your sterilizer. Compulsories and frequency of validation are regulated by local/national guidelines. more
      • Do I need to document the results of all periodic tests?
        W&H recommends recording all periodic tests, repairs and modifications on the sterilizer in a logbook including date and signature of the person who carried out the test or other activity. Compulsoriness of recording is regulated by local/national guidelines. more
      • How long should I archive the documentation of regular tests?
        Keep the results according to your country/regional specific laws and guidelines. Those records can be of assistance in case of service as well. more
      • Does my sterilizer need any periodic qualification/validation?
        Validation generally comprises a range of tests defined in internationally recognised standards. These tests begin for every device with the factory tests at W&H. In addition, corresponding documentation is included with your device. Prior to commissioning the device at your premises, the technician performs further checks to ensure the device functions correctly. Then, depending on the national/regional laws and directives, the device must be subjected to a so-called renewed performance qualification at regular intervals. For further information, please contact your dealer. more
      • Whom should I contact to validate and maintain my sterilizer?
        Validation shall be carried out by an official authority. Maintenance tasks shall be carried out by a qualified and authorized technician. more
      • How do I change consumable components on my sterilizer?
        Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available:
        • in the sterilizer, in the maintenance section of the menu; • in the Consumables & Accessories and Video Tutorial section of this web site. more
      • How do I clean the water tanks and what products should I use?
        Water tank cleaning instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available:
        • in the sterilizer, in the maintenance section of the menu;
        • in the Consumables & Accessories and Video Tutorial section of this web site. more
      • Should I pouch wrap my instruments and how long can I store them after sterilization?
        To ensure that your instruments are stored in a sterile state from their successful sterilization until they are used, W&H recommends wrapping them in suitable sterile barrier systems before sterilization. The recommended storage period can be found in the recommendations of the wrapping material manufacturer. For further information, please contact your dealer. more
      • The sterilizer remains switched OFF.
        POSSIBLE CAUSE
        The main switch or network circuit breaker is OFF.
        SOLUTION
        Activate the main switch or network circuit breaker (ON).

        POSSIBLE CAUSE
        No voltage at the socket.
        SOLUTION
        Check the building electrical supply.

        POSSIBLE CAUSE
        The power cord is not connected properly
        SOLUTION
        Check and connect the power cord properly more
      • Water is leaking at the front of the sterilizer.
        POSSIBLE CAUSE
        Leaks through the chamber door seal.
        SOLUTION
        Clean or replace the door seal. Clean the door surface facing the chamber.

        POSSIBLE CAUSE
        Internal leak.
        SOLUTIONS
        Contact an authorized W&H service partner. more
      • At the end of the cycle, there is residual water in the chamber
        POSSIBLE CAUSE
        Sterilizer not properly levelled
        SOLUTION
        Place the sterilizer on a flat, levelled surface.

        POSSIBLE CAUSE
        Overloaded chamber
        SOLUTION
        Comply with the maximum load weight limits for each type of load. Always use the chamber rack for trays and cassettes.

        POSSIBLE CAUSE
        Chamber filter clogged
        SOLUTION
        Remove and clean the chamber filter.

        POSSIBLE CAUSE
        Load incorrectly placed
        SOLUTION
        Follow the recommendations as listed in chapter “Load maintenance and preparation” in the Instructions for Use. more
      • The sterilizer enters into “Standby mode” immediately after being switched on.
        POSSIBLE CAUSE
        The unit was turned off while in standby mode.
        SOLUTION
        Press the “Standby mode” button to exit. more
      • When starting a cycle, the chamber door locks but re-opens immediately. The “Open the door” message appears.
        POSSIBLE CAUSE
        Door seal not properly placed.
        SOLUTION
        Ensure that the door seal is correctly inserted around the whole circumference.

        POSSIBLE CAUSE
        Door jammed by external objects or by the load itself
        SOLUTION
        Remove any objects interfering with the chamber door. Check for interference between the door and the chamber furniture and/or the load. more
      • Stains, spots or colour change on instruments.
        POSSIBLE CAUSE
        Instruments rinsed with tap water and not dried before sterilization.
        SOLUTION
        Ensure that instruments are rinsed with distilled or demineralized water and completely dry before they are placed in the sterilizer.

        POSSIBLE CAUSE
        Use of water of poor quality or water containing chemical substances
        SOLUTION
        Drain both water tanks. Use good quality water (see chapter “Water quality” in the Instructions for Use).

        POSSIBLE CAUSE
        Organic or chemical residues on the instruments
        SOLUTION
        Clean, rinse and dry instruments before placing them in the sterilizer (see chapter “Load maintenance and preparation” in the Instructions for Use).

        POSSIBLE CAUSE
        Dirt in chamber, trays, tray rack.
        SOLUTION
        Clean the chamber and wash the chamber furniture

        POSSIBLE CAUSE
        Contact between instruments of different materials
        SOLUTION
        Ensure that instruments of different materials do not touch each other (aluminum, carbon or stainless steel, etc.); place them on different trays or cassettes or pouch them (see chapter “Load maintenance and preparation” in the Instructions for Use).

        POSSIBLE CAUSE
        Scale deposits on the chamber
        SOLUTION
        Clean the chamber and use good quality water (see chapter “Water quality” in the Instructions for Use). more
      • No cycles are stored in the cycle history menu
        POSSIBLE CAUSE
        An electronic board was replaced when serviced.
        SOLUTION
        The memory of the old board cannot be restored. Consult the cycle history in the USB pen drive. more
      • Warning about saving to the USB (HTML and SCL files)
        POSSIBLE CAUSE
        The USB pen drive is disconnected or full.
        SOLUTION
        Check presence and condition of the USB pen drive. If the problem persist, contact an authorized W&H service partner. more
      • Warning about programmed maintenance.
        POSSIBLE CAUSE
        The maintenance program needs to be carried out.
        SOLUTION
        Contact an authorized W&H service partner. more
      • The cycle report printer does not work.
        POSSIBLE CAUSE
        Printer not properly connected or not powered
        SOLUTION
        Check the data and the power connection to the printer. more
      • The sterilizer is connected to an automated water supply system, there is no clean water in the tank and the automatic water filling does not fill the water.
        POSSIBLE CAUSE
        Water filling system not connected.
        SOLUTION
        Connect the water filling system to the sterilizer. (see Instructions For Use).

        POSSIBLE CAUSE
        When the water filling system attempted to fill the tank, water was temporarily unavailable.
        SOLUTION
        Since the water tank filling is attempted only once in-between cycle execution, this event inhibits water feeding. Switch the sterilizer OFF and then ON again.
        Check the external water supply system.
        Check for water leaks from the sterilizer.

        POSSIBLE CAUSE
        Faulty MIN water level sensor in the clean water tank
        SOLUTION
        Contact an authorized W&H service partner. more
      • The sterilization (PROCESS) phase of a sterilization cycle was longer than expected.
        POSSIBLE CAUSE
        The chamber temperature dropped below the minimum threshold and the software performed a successful recovery
        SOLUTION
        Wait for cycle completion. If the problem occurs frequently, contact an authorized W&H service partner. more
      • Warning about consumable replacement.
        POSSIBLE CAUSE
        A consumable needs to be replaced.
        SOLUTION
        Order the requested consumable (door seal and HEPA/bacteriological filter) and replace it. See the chapter “Maintenance” in the Instructions for Use. more
      • What is the maximum noise level of the Lexa Plus sterilizer?
        The max. noise level of the Lexa Plus is 70 dB. more
      • Which water quality should I use for my Lexa Plus sterilizer?
        The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardised. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Multidem system and OSMO are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use). more
      • Do I need to carry out periodic testing on my W&H Sterilizer?
        All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines. more

      Video Tutorials

      For a seamless and interactive online experience of our multimedia content (e.g. videos, maps), please accept the storage of the necessary cookies (YouTube, MovingImage - detailed information on these cookies can be found in the imprint/data protection) for marketing purposes.

      Accept
    • Lara sterilization cycles

      Sterilization cycle B-Universal 134 B-Prion 134 B-Universal 121 S FAST 134 (optional)
      Temperature: 134 °C 134 °C  121 °C 134 °C
      Holding time (minutes): 4 18.5 20.5 3.5
      Total cycle duration
      (minutes) including
      complete drying:
      Lara 17
      Automatic from
      25 to 51
      37
      Automatic from
      40 to 65
      52 41 to 73 20 to 22
      Lara 22
      Automatic from
      28 to 55
      39
      Automatic from
      40 to 68
      54 42 to 76 20 to 22
      Lara 17 loads:
      Lara 22 loads:
      maximum solid 4.5 kg/typical 2 kg/maximum porous 1.5 kg
      maximum solid 5.5 kg/typical 2 kg/maximum porous 2 kg
      max. unwrapped 2 kg
      max. unwrapped 2 kg
      Lara 17/22 containers load: maximum 9 kg maximum 9 kg
      Test cycles: Helix/Bowie Dick/Vacuum test

      Items supplied with: reversible rack, 3 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test  

      Lara

      Trade name: Lara
      Type: RIL-117 RIL-122
      Chamber size: 17 l 22 l
      Power supply: 200–240 V AC; 50/60 Hz; 10 A
      Power consumption: 2.0–2.4 kW
      Overall dimension
      (width x height x depth):
      465 x 452 x 646 mm
      Weight: 42.5 kg 44 kg
      Main / Used water tanks: 4.8/4.8 l
      Working range: 7 to 15 cycles
      Usable space in chamber
      (width x height x depth)¹:
      195 x 195 x 312 mm 195 x 195 x 400 mm
      Connection types: 2 USB ports (5 as option), integrated automatic water filling valve

      (1) Usable space is referred to standard rack configuration.

      Lara sterilizers were designed, certified and validated with the most stringent directives and standards:

      N°2017/745

      Medical Device Regulation

      IEC 61326-1

      Electromagnetic compatibility

      2014/68/EU

      Pressure equipment Directive

      IEC 61010-1

      Safety requirements sterilizer

      EN 13060

      Small steam sterilizer

      IEC 61010-2-040

      Specific requirements for steam sterilizer

      IEC 61770

      Electric appliances connected to the water mains

      2012/19/EU

      Waste Electrical and Electronic Equipment

      The sterilizer can be validated according to EN ISO 17665-1