Cleaning processThe efficient ONE
Cleaning the spray channels reduces potential blockages caused by limescale deposits and therefore ensures smoother operation, longer lifespan, and less service costs.
Lubrication processThe effective ONE
With the ground-breaking technology of oil nebulisation, dental transmission instruments are lubricated even faster and more thoroughly than ever before.
The oil is first atomised in the adaptor to a fine oil mist, which is then blown into the transmission instrument at high pressure. Debris from inside the transmission instrument is removed, and all gear parts are lubricated without putting them in motion.
With the active suction and HEPA filter, aerosols are removed from the chamber, keeping the working environment clean and safe.
The integrated ONE
Sterilization, hygiene, and maintenance are increasingly important in the dental practice. The new Assistina One is an optimally matching part of W&H´s wide range of products for the safe reprocessing of instruments, contra-angle and turbine handpieces according to the recommended, holistic W&H AIMS hygiene workflow.
W&H AIMS
W&H AIMS: A how-to workflow video for Advanced Infection Prevention Management
Product sets & consumables
Assistina One MB-301 + 1 Quick RM + 1 Quick ISO
REF 30398003Includes:
› Air connection hose (2m)
› One Care Set
› 1 Quick ISO adaptor
› 1 Quick RM adaptorOne Care Set MB-301
REF 08064970Includes:
› 1 cartridge of Activefluid (200 ml)
› 1 cartridge of Service Oil F1 (200 ml)
› 1 HEPA filterDownload Area
FAQs
-
What does HEPA mean, and what is filtered out of the air?
HEPA is short for 'High Efficiency Particular Air', and the filter is classified as HEPA class E11. This means that the filter retains all particles larger than 1 µm, which includes, among other things, a wide range of bacteria and viruses, suspended particles such as dust and respirable aerosols. more
-
Why there is no rotational lubrication option in the Assistina One?
The oil is atomized in the Assistina One before it enters the instrument. This ensures full coverage of the gear parts, eliminating the need for technically complex and time-consuming rotational lubrication. However, there is no negative impact on the maintenance quality. more
-
What is the difference between oil nebulization and rotational lubrication?
Final result is the same, but in a shorter time. In the nebulization system, oil is nebulized in the adaptor before it is blown into the handpiece. An oil mist coats all gear parts. more
-
What instrument adaptors are available for the Assistina One?
All adaptors that are suitable for the Assistina 3x2/3x3 can also be used in the Assistina One. There is only one exception: the RM/ISO adaptor for the Assistina 3x2/3x3 is not compatible with the Assistina One. Quick Connect adaptors ISO and RM can be used for the Assistina One instead. An overview of all adaptors is available in the product website. more
-
Why does the Assistina One have a HEPA filter?
During the maintenance process in the Assistina One, the rotating parts are lubricated with oil, and the spray channels rinsed with cleaning solution. These liquids are introduced to the handpieces using compressed air, which produces aerosols that may contain potentially contaminated body fluids (e.g. blood and saliva). In order to ensure that the apparatus poses no risk to the user, the patient or third parties, the legislation stipulates that aerosols must be prevented from leaking out of the apparatus. We make sure of this by using active suction and the HEPA filter. This HEPA filter must of course be changed regularly, which is why the lifespan of the filter is matched exactly to the capacity of the two cartridges. The HEPA filter therefore reaches the end of its service life once one set of cartridges has been used. more
-
How does the user know when the ONE Care Set needs to be replaced?
The ONE Care Set must be replaced when the cartridges are empty. The levels can be checked thanks to the device's fill level indicators, located on the right and left sides. more
-
How often does the ONE Care Set needs to be replaced?
The consumption volumes for each instrument are 0.07 ml of oil and 0.07 ml of cleaning solution. This gives a maximum capacity of 2857 cycles. The value is around 2800 to 2850 cycles, as initial filling is also required and the fill quantity of the cartridges is subject to certain tolerances that are regulated by the applicable standards. more
-
How do you replace the ONE Care Set?
The two cartridges and the filter can be changed easily without any tools. The fluid is supplied via tapping needles in the bottom of the cartridge spaces that penetrate the cartridge membrane, thereby drawing the fluid out of the cartridge. The two cartridges have different shapes to ensure that the two fluid types cannot be mixed up. In addition, it is only possible to install the HEPA filter incorrectly by using force, which is not necessary. more
Instructions for use
Find all languages of the Update Hygiene and Maintenance (EN ISO 17664:2017) in the Download Centre.
Video Tutorials
-
What does HEPA mean, and what is filtered out of the air?
-
System overview adaptor
Assistina One (MB-301) with Quick Connect
Step 1: Select the handpiece type
- Straight and contra-angle handpieces: require Quick ISO adaptor
- Turbine handpieces: require Quick RM adaptor
Step 2: For brand specific connections
- For brand specific connections, like Roto Quick, select the right adaptor suitable for you.
MCA-QC-ISO adaptor Quick ISO
REF 06422600
MCA-QC-RM adaptor Quick RM
REF 06395100
- Straight and contra-angle handpieces: require Quick ISO adaptor
-
Technical Data
Type Assistina One
MB-301Process: Cleaning the spray channels, lubrication Suitable for: Straight and contra-angle handpieces, turbine handpieces, air scalers, air motors Air consumption: approx. 50 Nl/min. Operating pressure: 5–10 bar (controlled via an integrated, automatic pressure controller)
Dimensions (W x D x H): 166 × 383 × 243 mm Weight: 2.2 kg Capacity: 200 ml W&H Activefluid cleaning solution
200 ml W&H Service Oil F1