FAQ - Întrebări frecvente
Întrebări frecvente despre produsele W&H și descrieri utile pe teme de igienă și îngrijire.
Cuple
-
Roto Quick (5)
-
Can the W&H coupling be thermodisinfected and sterilized?
No, thermodisinfection of the coupling is not permitted; sterilization in steam sterilizers can be applied at a temperature of 135°C. Hot air sterilization and/or cold sterilization (placing in liquids) are not approved.
-
Are there other couplings than the Roto Quick available?
Yes. W&H also offers the "RM-34 LED" for Multiflex® connection.
-
Can I use all instruments which fitted the old Roto Quick coupling (924, ...) with the new RQ-24 too?
Yes. All existing and new instruments can be used with couplings from both the old and the new generations.
-
How does the new W&H Click & Pull system work? How do I remove a dental turbine?
Use your thumb and index finger to retract the coupling's retention sleeve. The turbine can now be removed from the coupling without any problems. Pulling forcefully on the dental turbine can cause damage. If excessive force is required to remove the turbine, verify that the retention sleeve on the coupling can be moved without any problems.
-
Why is a 6-hole connection required for the use of an LED turbine?
A 6-hole connection for the turbine tubing is necessary to supply the turbine with electricity via electrical contacts. Other connections do not provide a power supply.
-
Can the W&H coupling be thermodisinfected and sterilized?
Accesorii
-
Anse detartraj Piezo (5)
-
Are W&H tips (e.g. 1U) compatible with the new W&H high-speed mounting system?
No, they are not compatible because they have different threads.
-
How do I know which tip goes with which handpiece?
There are symbols on the tip, the tip changer and the handpiece identifying which system they belong to.
-
Are the new handpieces compatible with the PA-123/PA-115?
No, they are not compatible.
-
Can the tips and tip changers be prepared in the sterilizer?
Yes, as long as your handpiece or contra-angle handpiece carries the mark “sterilizable up to the stated temperature”.
-
Are the tips and tip changers thermo washer disinfectable?
Yes, as long as your handpiece or contra-angle handpiece carries the mark “thermo washer disinfectable”.
-
Are W&H tips (e.g. 1U) compatible with the new W&H high-speed mounting system?
-
Periuțe și cupe profilaxie (3)
-
Can I use the W&H Proxeo Twist LatchShort prophy cups in a contra-angle handpiece with a standard 2.35-mm chuck?
Due to the shorter shaft length of the W&H prophy cups, they cannot be gripped by a standard contra-angle handpiece chuck, so it is not possible to drive the W&H prophy cup.
-
Can I use the Proxeo Twist LatchShort Prophy cups and brushes multiple times?
No, the Proxeo Twist Prophy cups and brushes are single-use only.
-
Can I use my LatchShort Prophy cups and brushes with other contra-angle handpieces?
No, LatchShort Prophy cups and brushes are designed to be used only with the WP-66 W contra-angle handpiece.
-
Can I use the W&H Proxeo Twist LatchShort prophy cups in a contra-angle handpiece with a standard 2.35-mm chuck?
-
Pudră profilaxie (2)
-
Can I use any polishing powder available on the market in the Proxeo Aura air polishing handpiece?
The Proxeo Aura air polishing handpiece and the relevant powders were developed as a complete treatment system, in order to achieve the best possible treatment results and to ensure the reliability of the system. The use of other powders is not approved. This can lead to increased abrasiveness and risk of blockage or reduced efficiency.
-
Can I use the W&H prophylaxis powder for other air polishing handpieces?
This depends on the manufacturer’s specifications for the relevant air polishing handpiece/device.
-
Can I use any polishing powder available on the market in the Proxeo Aura air polishing handpiece?
Dispozitive de curățare și dezinfecție
-
Thed+/Thed (4)
-
Dacă folosesc un termodezinfector pentru reprocesarea instrumentelor, există riscul să le deteriorez și să fie necesară înlocuirea lor mai frecventă?
Categoric, nu. Dimpotrivă, situația este exact opusă. Prin utilizarea unui dispozitiv automat, cum ar fi termodezinfectorul, se asigură cel mai înalt standard posibil de curățare a instrumentelor, validat și previzibil, fără riscul de erori datorate intervenției operatorului, optimizându-se astfel durata de viață a instrumentelor.
-
Pot utiliza accesorii, cum ar fi coșuri, suporturi etc., de la vechiul meu termodezinfector?
Toți producătorii validează eficiența de spălare a mașinilor lor în combinație cu accesorii concepute în mod special pentru acestea. Prin urmare, nici succesul final, nici validarea nu sunt garantate atunci când se utilizează accesorii care nu au fost create și validate de producător.
-
Din moment ce folosesc deja un sterilizator în cabinetul meu, de ce ar trebui să folosesc și un termodezinfector?
Termodezinfectoarele și autoclavele sunt echipamente care nu pot lipsi dintr-un flux de lucru modern, având în vedere că îndeplinesc sarcini diferite, dar în același timp absolut complementare. Obiectivul principal al unui termodezinfector este de a curăța toate instrumentele dentare în mod riguros și validat, protejând operatorii de riscul de infecții încrucișate și reducând încărcătura patogenă, datorită procesului de termodezinfecție. Toate aceste faze pregătitoare sunt absolut necesare pentru a se asigura că etapa ulterioară, de sterilizare, poate avea loc în condiții ideale.
-
De ce, după reprocesarea instrumentelor în termodezinfector, am observat oxidarea sau decolorarea suprafețelor acestora?
Fenomenele de oxidare și/sau decolorare nu pot fi atribuite, aproape niciodată, în mod direct activității termodezinfectorului; în majoritatea cazurilor, acestea pot fi explicate prin interacțiunile dintre reziduurile produselor chimice prezente pe instrumente, provenite eventual din procesul anterior de decontaminare într-un rezervor static, și substanțele chimice utilizate în procesele termodezinfectorului. În acest sens, este întotdeauna recomandabil să se clătească bine instrumentele înainte de a le încărca în termodezinfector.
-
Dacă folosesc un termodezinfector pentru reprocesarea instrumentelor, există riscul să le deteriorez și să fie necesară înlocuirea lor mai frecventă?
-
U-Son (4)
-
Există instrumente care nu pot fi reprocesate într-un aparat de curățare cu ultrasunete?
Piesele de mână nu trebuie niciodată reprocesate într-un echipament cu ultrasunete, pentru a evita deteriorarea mecanismelor interne. În general, este recomandat ca întodeauna să se respecte atent și cu strictețe instrucțiunile de întreținere a pieselor de mână, comunicate de producător.
-
Pot utiliza aparatul de curățat cu ultrasunete fără coșul pentru instrumente?
Nu este posibilă utilizarea dispozitivului fără coșul pentru instrumente, având în vedere că instrumentele nu trebuie să ajungă niciodată în contact direct cu baza cuvei, pentru a nu interfera cu emisia de ultrasunete de la transductoare.
-
Cât de des trebuie să schimb soluția de curățare?
Soluția de curățare utilizată în aparatul de curățat cu ultrasunete trebuie înlocuită cel puțin o dată pe zi (sau chiar mai frecvent, dacă acest lucru este indicat de producătorul substanțelor chimice), sau ori de câte ori se observă prezența evidentă a unor depuneri de reziduuri la fundul cuvei.
-
Cum pot evalua dacă aparatul meu de curățat cu ultrasunete funcționează bine?
Există pe piață mai multe tipuri de indicatori pentru verificarea eficacității unui dispozitiv cu ultrasunete, însă testul cu folie de aluminiu rămâne, chiar și astăzi, printre cele mai utilizate, datorită raportului cost-eficacitate, a caracterului practic, a vitezei și a fiabilității sale.
-
Există instrumente care nu pot fi reprocesate într-un aparat de curățare cu ultrasunete?
Piese Chirurgicale Fierăstrău
-
How to dismantle the handpiece
See detailed description in user manual or Video.
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
Măsurarea stabilității implantului
-
Osstell Classic (3)
-
Where can I buy the SmartPegs?
Please contact your local W&H partner
The sale of SmartPegs will be defined per country
The surgeon will receive 5 free SmartPegs. A voucher will be provided with the delivery of the product. -
How does the measurement of the ISQ work?
> To measure the ISQ of an implant, you first screw a so called SmartPeg into the placed implant. The SmartPeg, with its magnet on top, works like a small tuning fork. The magnet on the SmartPeg is ‘hit’ with magnetic pulses, from the probe, which makes the SmartPeg vibrate. Due to the stiffness in the interface between the implant surface and the bone the SmartPeg will vibrate accordingly. The more dense the bone is the higher stability and the higher ISQ value.
-
What does ISQ stand for?
The abbreviation ISQ stands for Implant Stability Quotient. It is a value between 1 and 100 and gives the surgeon an insight of the implant stability after placing the implant. With the ISQ value the surgeon can monitor the osseointegration.
-
Where can I buy the SmartPegs?
-
Osstell Beacon (4)
-
Este coeficientul ISQ dovedit științific?
Tehnologia Osstell ISQ este dovedită de peste 1400 de studii științifice. Aceste studii au arătat că există corelare între cuplu și ISQ.
-
Where can I buy the SmartPegs?
Please contact your local W&H partner
The sale of SmartPegs will be defined per country
The surgeon will receive 5 free SmartPegs. A voucher will be provided with the delivery of the product. -
How does the measurement of the ISQ work?
> To measure the ISQ of an implant, you first screw a so called SmartPeg into the placed implant. The SmartPeg, with its magnet on top, works like a small tuning fork. The magnet on the SmartPeg is ‘hit’ with magnetic pulses, from the probe, which makes the SmartPeg vibrate. Due to the stiffness in the interface between the implant surface and the bone the SmartPeg will vibrate accordingly. The more dense the bone is the higher stability and the higher ISQ value.
-
What does ISQ stand for?
The abbreviation ISQ stands for Implant Stability Quotient. It is a value between 1 and 100 and gives the surgeon an insight of the implant stability after placing the implant. With the ISQ value the surgeon can monitor the osseointegration.
-
Este coeficientul ISQ dovedit științific?
Detartraj pneumatic
-
Proxeo (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
Micromotoare Electrice
-
Micromotoare Electrice (EM-E6) (2)
-
I've got a micromotor without light. Do I have to change the micromotor and the contra-angle to have light?
No. You can purchase an Allegra LED G contra-angle which has its own light generator and a LED. No further investment necessary.
-
Does the electric motor need to be greased?
No, the motor bearings require no maintenance. Any additional greasing would reduce the service life of the bearings.
-
I've got a micromotor without light. Do I have to change the micromotor and the contra-angle to have light?
-
Set Upgrade LED (2)
-
I've got a micromotor without light. Do I have to change the micromotor and the contra-angle to have light?
No. You can purchase an Allegra LED G contra-angle which has its own light generator and a LED. No further investment necessary.
-
Does the electric motor need to be greased?
No, the motor bearings require no maintenance. Any additional greasing would reduce the service life of the bearings.
-
I've got a micromotor without light. Do I have to change the micromotor and the contra-angle to have light?
Agenți Curățare & Dezinfectare
-
BePro (5)
-
Care sunt principalele aspecte pe care trebuie să le luăm în considerare atunci când alegem un dezinfectant?
Există trei aspecte fundamentale pe care trebuie să le luăm întotdeauna în considerare, atunci când alegem un dezinfectant: eficacitatea dorită, adică abilitatea produsului de a fi activ asupra unei game largi de agenți patogeni, eficiența așteptată, adică abilitatea produsului de a atinge eficacitatea maximă în cel mai scurt timp posibil și, în cazul produselor concentrate, cu cea mai mică concentrație posibilă și, în cele din urmă, compatibilitatea necesară, adică posibilitatea de a utiliza produsul fără riscul de a deteriora, distruge sau compromite utilizarea instrumentului sau a suprafeței care urmează să fie tratată.
-
Sunt toți dezinfectanții la fel?
Dezinfectanții nu sunt toți la fel, existând diferite tipuri de dezinfectanți, în funcție de formulele chimice ale acestora și de principiul (principiile) activ(e) conținut(e).
-
Cum aleg dezinfectantul potrivit pentru cabinetul meu?
În funcție de categoriile de microorganisme patogene, asupra cărora este eficient dezinfectantul, se disting diferite niveluri de dezinfectare: ridicat, intermediar și scăzut. Pentru a asigura o siguranță totală în cabinet, atât pentru pacienți, cât și pentru personalul operator, alegerea dezinfectantului ce urmează a fi utilizat, trebuie să se îndrepte întotdeauna spre un produs al cărui nivel de dezinfectare să fie ridicat.
-
Sunt toți agenții patogeni la fel de rezistenți la dezinfectanți?
Nu, agenții patogeni au rezistențe foarte diferite, în funcție de substanța dezinfectantă folosită în încercarea de a-i elimina. Agenții patogeni, cel mai puțin rezistenți, variază de la virusurile încapsulate până la micobacterii, care sunt în schimb cei mai rezistenți și necesită un nivel ridicat de dezinfectare pentru a-i elimina.
-
Există agenți patogeni mai rezistenți decât micobacteriile?
Da, agenții patogeni cel mai dificil de eliminat sunt sporii și prionii, pentru a căror eliminare este necesar să se recurgă la utilizarea autoclavului pentru sterilizare.
-
Care sunt principalele aspecte pe care trebuie să le luăm în considerare atunci când alegem un dezinfectant?
Accesorii
-
Ulei Service F2 (6)
-
Can I use any oil to lubricate my W&H instruments?
W&H recommends that W&H instruments are lubricated with W&H oil which has been designed and developed to enhance the performance of your W&H instrument.
-
Why is W&H oil better than other lubricants?
Extensive research and testing produced a formula which best covers the extreme demands of dental instruments. It consists of an all-synthetic special oil of the highest purity supplemented by the specially developed additive package.
-
Up to what maximum temperature is the oil stable without problems?
W&H Service Oil F1 has longterm stability between –30 oC and +160 oC.
-
Can W&H Service Oil F1 be sterilized?
Yes, W&H has checked the sterilizability precisely. W&H Service Oil F1 was not found to influence the sterilization of the instruments in any way.
-
Which aerosol is used in the W&H F1 spray can and what properties does it have?
A mixture of propane and butane (standard aerosol in all spray cans). It contains no CFCs and as such is environmentally-friendly.
-
How should I dispose of empty cans?
In accordance with the national regulations for aerosol cans. More information on disposal can be found on the safety datasheet.
-
Can I use any oil to lubricate my W&H instruments?
-
Seal2 (1)
-
Ar trebui să pun în pungă/împachetez instrumentele și cât timp pot să le păstrez după sterilizare?
Pentru a vă asigura că instrumentele dvs. sunt păstrate în stare sterilă de la sterilizarea lor până la utilizare, W&H recomandă să le împachetați, utilizând sisteme de bariere sterile adecvate (pungi/role de sterilizare) înainte de sterilizare. Perioada de depozitare recomandată poate fi găsită în recomandările producătorului materialului de împachetare. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Ar trebui să pun în pungă/împachetez instrumentele și cât timp pot să le păstrez după sterilizare?
Piese de mână drepte & contraunghi
-
Synea (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
-
Alegra (9)
-
Can the new Alegra instruments be connected to motors with a light supply?
Yes, the straight and contra-angle handpieces of the new Alegra range can be used on any motor with an ISO connection. .
-
De ce piesa contraunghi WE-99 are raportul de transmisie de 1:4,5?
În acest fel, cu ajutorul piesei WE-99 se va atinge, cu multă ușurință și precizie, viteza optimă a frezelor rapide.
-
Can LED instruments with generator be sterilized?
Yes, W&H handpieces with LED and generator are both sterilizable
and thermodisinfectable
.
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
For what reason may the LED fail to light up?
Because the energy for the LED light in the new Alegra instruments is generated by an integrated generator, the brightness of the LED depends on the motor speed (at least 9.000rpm).
-
What are the advantages of the new Alegra instruments?
Compatible LED light – the instruments in the new Alegra range have LED light, which is supplied with energy by an integrated generator. The instruments are therefore independent of the light supply of the motor and the dental unit and offer the possibility of working with light compatible with any unit. The technology of the integrated generator means that there is only illumination when the instrument is running.
-
Can the HE-43 E handpiece be employed in surgical applications?
The handpieces in the new Alegra range have been tested and approved for the following applications removing cariogenic materials, preparing cavities and crowns, removing fillings and finishing tooth and restoration surfaces. The use of the handpieces in other application areas is not permitted and the user bears all responsibility in cases where this instruction is disregarded.
-
Is more heat generated by the light source when LED technology is employed?
No, because they are usually operated with coolant air. In addition, the LED contra-angle handpieces are equipped with an internal spray function, which guarantees additional cooling.
-
What is the difference between the Alegra contra-angle handpieces WE-56 T and WE-56?
Alegra contra-angle handpieces designated with the letter "T" are significantly lighter in weight. Alegra contra-angle handpieces that are not designated with the letter "T" have been approved for automated thermo washer disinfection.
-
Can the new Alegra instruments be connected to motors with a light supply?
Piese de mână drepte & contraunghi
-
Endea (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
-
Endo NiTi (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
-
Endea Endo Cursor (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
Piese de mână drepte & contraunghi
-
Piese contraunghi (10)
-
Which material is used at W&H surgical instruments?
Internal and external parts are of high quality stainless steel (including the contra-angle handpiece head).
-
Is the removable spray clip of my surgical hand piece sterilizable?
Yes. In order to extend the live span of the clip, W&H recommends to remove the clip from the hand piece and sterilize it separately (hand piece and clip in one pouch but separated).
-
What is the difference between WI-75 and WS-75?
Both assure maximum performance. The WS-75 can be dismantled and therefor it can be cleaned more thoroughly.
-
Can LED instruments with generator be sterilized?
Yes, W&H handpieces with LED and generator are both sterilizable
and thermodisinfectable
.
-
How to dismantle the handpiece
See detailed description in user manual or Video.
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Why doesn't the light on my LED G contra-angle handpiece work when I am positioning an implant?
The generator in the contra-angle handpiece requires at least 300 rpm to produce enough electricity for the LED. The rotary speed is between 15 and 50 rpm when positioning an implant.
-
Se pot atașa piesele contraunghi actuale la micromotoare?
Da. Toate piesele drepte și contraunghi de chirugie W&H pot fi atașate la micromotorul EM-19.
-
Se pot atașa piesele contraunghi la micromotoarele vechi?
Nu. Pentru a asigura un echilibru optim, am scurtat lungimea cuplei interioare. Prin urmare, piesele contraunghi cu sufixul „L” nu pot fi atașate la micromotoarele mai vechi, ci numai la EM-19 și EM-19 LC.
-
Pot clienții noștri să demonteze piesele contraunghi?
Da. La piesele W&H, litera „S” indică întotdeauna că respectiva piesă cot chirurgicală poate fi demontată. Litera „I” indică faptul că această piesă nu poate fi demontată (de exemplu, WS-75 L vs. WI-75).
-
Which material is used at W&H surgical instruments?
-
Piese drepte (9)
-
Which material is used at W&H surgical instruments?
Internal and external parts are of high quality stainless steel (including the contra-angle handpiece head).
-
Care este lungimea minimă a mandrinei frezei pentru piesa S-11?
Mandrina frezei trebuie să aibă o lungime de cel puțin 44 mm („mandrină ISO 104”).
-
Can LED instruments with generator be sterilized?
Yes, W&H handpieces with LED and generator are both sterilizable
and thermodisinfectable
.
-
What is the minimum shaft lenght of the bur for the S-12?
The shaft of the bur should be at least 53mm long.
-
How to dismantle the handpiece
See detailed description in user manual or Video.
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Se pot atașa piesele contraunghi actuale la micromotoare?
Da. Toate piesele drepte și contraunghi de chirugie W&H pot fi atașate la micromotorul EM-19.
-
Se pot atașa piesele contraunghi la micromotoarele vechi?
Nu. Pentru a asigura un echilibru optim, am scurtat lungimea cuplei interioare. Prin urmare, piesele contraunghi cu sufixul „L” nu pot fi atașate la micromotoarele mai vechi, ci numai la EM-19 și EM-19 LC.
-
Pot clienții noștri să demonteze piesele contraunghi?
Da. La piesele W&H, litera „S” indică întotdeauna că respectiva piesă cot chirurgicală poate fi demontată. Litera „I” indică faptul că această piesă nu poate fi demontată (de exemplu, WS-75 L vs. WI-75).
-
Which material is used at W&H surgical instruments?
Echipamente Chirurgie
-
Implantmed (24)
-
Can my Implantmed motor be processed via thermal washer disinfector?
Yes, if the thermal washer disinfector has a drying cycle and the motor bears the sign that it is "Thermo washer disinfectable".
-
Am nevoie de un capac de protecție pentru motorul Implantmed?
Nu este necesar niciun capac de protecție.
-
Does my W&H unit need to be serviced.
Regular servicing of function and safety including the accessories is necessary and should be carried out at least once every three years, unless shorter intervals are prescribed by law. Detailed information can be found in the user manual at chapter "Servicing".
-
Ce tipuri de tuburi pot folosi cu dispozitivul meu?
Pentru micromotoarele cu lungimea de 1,8 m puteți utiliza tuburi de unică folosință (04363600) sau tuburi sterilizabile (0417XXX). Pentru combinația dintre modulul Implantmed și modulul Piezomed, oferim un tub cu valvă în Y (08072750). De asemenea, puteți găsi tuburile adecvate în instrucțiunile de utilizare ale dispozitivului.
-
How can the unit / motor be cleaned / sterilized?
Please find a detailed description in the user manual at chapter "Hygiene and Maintenance".
-
Can handpieces from another manufacturer be used on a W&H surgical motor (implantmed)?
For applications which do not require accurate torque (drilling, grinding,…), any handpiece with ISO 3964 connection can be used. Please be informed that hand pieces from other manufacturers might not be able to handle the torque of your unit/motor and therefore might get damaged.
-
Se pot utiliza și alte mărci de tuburi?
Nu, sunt permise doar tuburile de irigare autorizate de W&H. Acest lucru se datorează specificațiilor (material, dimensiuni, debit de lichid de răcire), care ar putea cauza probleme precum scurgeri, răcire insuficientă sau chiar deteriorarea dispozitivului.
-
How to insert irrigation tube?
Please find a detailed description in the user manual at chapter "Starting operation - General".
-
Este posibilă reglarea vitezei micromotorului în cadrul programului de inserare a implantului?
Da, atât cuplul, cât și viteza pot fi setate pentru inserarea implantului.
-
Why doesn't the light on my LED G contra-angle handpiece work when I am positioning an implant?
The generator in the contra-angle handpiece requires at least 300 rpm to produce enough electricity for the LED. The rotary speed is between 15 and 50 rpm when positioning an implant.
-
Does the motor need to be greased?
No, the Implantmed/Elcomed motor does not need greasing. The motor bearings require no maintenance. Any additional greasing would reduce the service life of the bearings.
-
Se pot echipa dispozitivele Implantmed cu pedaala fără fir?
Dispozitivele Implantmed Plus pot fi modernizate. Pur și simplu conectați cheița dongle furnizată cu echipamentul și acționați-l folosind pedala fără fir. Pentru echipamentele Implantmed Plus II, trebuie să decideți dacă doriți o versiune cu sau fără cablu, la momentul achiziționării. Acest lucru nu poate fi modificat ulterior.
-
Pedalele vechi, cu cablu, sunt compatibile cu noul aparat sau invers?
Vezi tabelul general.
Pedală Implantmed Classic SI-9xx /Implantmed Plus SI-10xx Implantmed Plus II SI-2102 Implantmed Plus II SI-2101 Implantmed Plus II SI-2100 Pedală fără fir S-NW3 nu da da nu Pedală cu fir S-N3 nu nu nu da Pedale S-N2 / S-NW da nu nu nu -
Se pot atașa piesele contraunghi actuale la micromotoare?
Da. Toate piesele drepte și contraunghi de chirugie W&H pot fi atașate la micromotorul EM-19.
-
Se pot atașa piesele contraunghi la micromotoarele vechi?
Nu. Pentru a asigura un echilibru optim, am scurtat lungimea cuplei interioare. Prin urmare, piesele contraunghi cu sufixul „L” nu pot fi atașate la micromotoarele mai vechi, ci numai la EM-19 și EM-19 LC.
-
Pot clienții noștri să demonteze piesele contraunghi?
Da. La piesele W&H, litera „S” indică întotdeauna că respectiva piesă cot chirurgicală poate fi demontată. Litera „I” indică faptul că această piesă nu poate fi demontată (de exemplu, WS-75 L vs. WI-75).
-
What’s the difference between torque and ISQ?
Torque is a one time, static measurement at the time of placement of the implant and cannot be repeated later on in a non-invasive way. Osstell ISQ measures the lateral stability of the implant and the measurements can be repeated in a non-invasive, dynamic way to monitor the development of osseointegration.
-
Toate echipamentele Implantmed sunt compatibile cu modulul W&H Osstell ISQ?
Unitățile Implantmed Plus SI-1023, SI-1015 și SI-1010 pot fi actualizate cu modulul W&H Osstell ISQ. Pentru noua generație Implantmed Plus II se poate conecta Osstell Beacon.
-
Care sunt micromotoarele compatibile cu dispozitivele Implantmed?
Vezi tabelul general.
Micromotor Implantmed ClassicSI-9xx Implantmed PlusSI-10xx Implantmed Plus IISI-21xx EM-19 da da nu EM-19 LC nu da da Versiuni mai vechi ale micromotorului nu nu nu -
Ce înseamnă LC în EM-19 LC, există un LED în micromotor?
Particula LC înseamnă contacte pentru lumină. Aceasta înseamnă că am integrat o conexiune electrică în noile noastre micromotoare. Contactele pentru lumină alimentează cu energie piesele noastre de mână contraunghi cu sufixul „L” (de exemplu, WS-75 L), asigurând astfel o lumină optimă la orice viteză.
-
Este posibilă utilizarea Implantmed Classic cu pedala fără fir?
Nu, Implatmed Classic poate fi operat numai prin intermediul unei pedale cu fir.
-
Modulul Piezomed: se pot moderniza dispozitivele existente pe piață?
Da, dispozitivele pot fi modernizate de către utilizatori. Sunt posibile următoarele combinații: Implantmed Plus cu modulul Piezomed Plus sau modulul Piezomed Classic. Implantmed Plus II cu modulul Piezomed Plus II.
-
Modulul Piezomed: sunt piesele de mână ale modulelor SA-40 și SA-40 L compatibile cu dispozitivul Piezomed SA-320?
Nu, nu sunt compatibile.
-
Piezomed module: Does the module also work with OEM devices?
Yes, but different software updates are necessary.
-
Can my Implantmed motor be processed via thermal washer disinfector?
-
Piezomed (13)
-
How long do the instruments last and when should they be replaced?
The useful service life of the instruments is determined by the length of time they are used reprocessing and sterilization cycles have only marginal influence.
a. Instruments need to be replaced
- If there is a loss of power while working
- Where there is visible damage on the working part or the shank of the instrument.
b. Saws must be replaced if
- Teeth are broken off
- Teeth are worn
c. Diamond-coated instruments must be replaced
- As soon as the diamond coating is worn
-
Pot fi utilizate anse de alți producători în modulele Piezomed?
Nu
a. Filetul este diferit
b. Funcția de detectarea a anselor Piezomed nu poate detecta anse de altă marcă
-
How are the instruments prepared for cleaning?
a. Preparation in the ultrasonic bath:
- Place the instruments in the instrument tray and lower into the ultrasonic bath.
- Use cleaning agents and disinfectants suitable for hand instruments (probes, mirrors, forceps, etc.).
DO NOT USE a bur bath, as this is very aggressive.
- After cleaning in the ultrasonic bath, rinse adequately with water to remove any remnants of the cleaning agent and disinfectant from the coolant channels.
- Blow dry with compressed air after rinsing
- Replace dried instruments in the instrument tray and package for sterilization
b. Mechanical preparation in the cleaning unit and disinfector
- Use the cleaning adapter for cleaning units and disinfectors (see Instructions for Use – Accessories)
- After preparation in the cleaning unit and disinfector, check that the coolant channels are dry and, if necessary, dry again with compressed air. -
Can Piezomed instruments also be used on third-party devices?
No
a. Different thread
b. Third-party device cannot find the correct resonance point – instrument fails to oscillate -
Is it necessary for the torque driver to engage audibly when tightening the instrument?
Yes, it must engage audibly once, as otherwise the required torque is not achieved.
-
How long is the useful service life of the LED?
The user must replace the LED socket as soon as the protective coating on the LED discolours and the luminous power is impaired as a result. The length of time this takes can differ depending on the type of steriliser or sterilization process.
-
How long is the useful service life of the instrument detection?
The instrument detection is located in the LED socket and is automatically replaced at the same time as the LED socket.
-
Modulul Piezomed: se pot moderniza dispozitivele existente pe piață?
Da, dispozitivele pot fi modernizate de către utilizatori. Sunt posibile următoarele combinații: Implantmed Plus cu modulul Piezomed Plus sau modulul Piezomed Classic. Implantmed Plus II cu modulul Piezomed Plus II.
-
Modulul Piezomed: sunt piesele de mână ale modulelor SA-40 și SA-40 L compatibile cu dispozitivul Piezomed SA-320?
Nu, nu sunt compatibile.
-
Piezomed module: Does the module also work with OEM devices?
Yes, but different software updates are necessary.
-
Modulul Piezomed: Care sunt principalele diferențe între modulul Piezomed Plus
și modulul Piezomed Classic?
Caracteristici Modul Piezomed Plus Modul Piezomed Classic Putere 24 W 18 W Lumină LED fără lumină Detectare ansă da nu Culoare negru alb Lungime cablu 1.8 m și 3.5 m 1.8 m -
Piezomed module: Compatibility of the handpieces with Piezomed module Plus II, Piezomed module Plus
and Piezomed Classic module.
Handpieces Piezomed module Plus II Piezomed module Plus Piezomed Classic module SA-40 yes yes yes SA-40 L (light version) yes yes no SA-320 no no no -
Se poate utiliza piesa SA-40 împreună cu modulul Piezomed Module Plus II?
Da, dar funcția de detectare automată a ansei trebuie dezactivată din submeniu. Setările, cum sunt grupa de putere, puterea și fluxul de răcire, trebuie configurate de utilizator conform fișei tehnice a ansei (inclusă cu fiecare ansă piezoelectrică W&H).
-
How long do the instruments last and when should they be replaced?
-
Implantmed Plus II (28)
-
Ce tipuri de tuburi pot folosi cu dispozitivul meu?
Pentru micromotoarele cu lungimea de 1,8 m puteți utiliza tuburi de unică folosință (04363600) sau tuburi sterilizabile (0417XXX). Pentru combinația dintre modulul Implantmed și modulul Piezomed, oferim un tub cu valvă în Y (08072750). De asemenea, puteți găsi tuburile adecvate în instrucțiunile de utilizare ale dispozitivului.
-
Se pot utiliza și alte mărci de tuburi?
Nu, sunt permise doar tuburile de irigare autorizate de W&H. Acest lucru se datorează specificațiilor (material, dimensiuni, debit de lichid de răcire), care ar putea cauza probleme precum scurgeri, răcire insuficientă sau chiar deteriorarea dispozitivului.
-
Este posibilă reglarea vitezei micromotorului în cadrul programului de inserare a implantului?
Da, atât cuplul, cât și viteza pot fi setate pentru inserarea implantului.
-
Se pot echipa dispozitivele Implantmed cu pedaala fără fir?
Dispozitivele Implantmed Plus pot fi modernizate. Pur și simplu conectați cheița dongle furnizată cu echipamentul și acționați-l folosind pedala fără fir. Pentru echipamentele Implantmed Plus II, trebuie să decideți dacă doriți o versiune cu sau fără cablu, la momentul achiziționării. Acest lucru nu poate fi modificat ulterior.
-
Pedalele vechi, cu cablu, sunt compatibile cu noul aparat sau invers?
Vezi tabelul general.
Pedală Implantmed Classic SI-9xx /Implantmed Plus SI-10xx Implantmed Plus II SI-2102 Implantmed Plus II SI-2101 Implantmed Plus II SI-2100 Pedală fără fir S-NW3 nu da da nu Pedală cu fir S-N3 nu nu nu da Pedale S-N2 / S-NW da nu nu nu -
Se pot atașa piesele contraunghi actuale la micromotoare?
Da. Toate piesele drepte și contraunghi de chirugie W&H pot fi atașate la micromotorul EM-19.
-
Se pot atașa piesele contraunghi la micromotoarele vechi?
Nu. Pentru a asigura un echilibru optim, am scurtat lungimea cuplei interioare. Prin urmare, piesele contraunghi cu sufixul „L” nu pot fi atașate la micromotoarele mai vechi, ci numai la EM-19 și EM-19 LC.
-
Pot clienții noștri să demonteze piesele contraunghi?
Da. La piesele W&H, litera „S” indică întotdeauna că respectiva piesă cot chirurgicală poate fi demontată. Litera „I” indică faptul că această piesă nu poate fi demontată (de exemplu, WS-75 L vs. WI-75).
-
Toate echipamentele Implantmed sunt compatibile cu modulul W&H Osstell ISQ?
Unitățile Implantmed Plus SI-1023, SI-1015 și SI-1010 pot fi actualizate cu modulul W&H Osstell ISQ. Pentru noua generație Implantmed Plus II se poate conecta Osstell Beacon.
-
Care sunt micromotoarele compatibile cu dispozitivele Implantmed?
Vezi tabelul general.
Micromotor Implantmed ClassicSI-9xx Implantmed PlusSI-10xx Implantmed Plus IISI-21xx EM-19 da da nu EM-19 LC nu da da Versiuni mai vechi ale micromotorului nu nu nu -
Ce înseamnă LC în EM-19 LC, există un LED în micromotor?
Particula LC înseamnă contacte pentru lumină. Aceasta înseamnă că am integrat o conexiune electrică în noile noastre micromotoare. Contactele pentru lumină alimentează cu energie piesele noastre de mână contraunghi cu sufixul „L” (de exemplu, WS-75 L), asigurând astfel o lumină optimă la orice viteză.
-
Este posibilă utilizarea Implantmed Classic cu pedala fără fir?
Nu, Implatmed Classic poate fi operat numai prin intermediul unei pedale cu fir.
-
Modulul Piezomed: se pot moderniza dispozitivele existente pe piață?
Da, dispozitivele pot fi modernizate de către utilizatori. Sunt posibile următoarele combinații: Implantmed Plus cu modulul Piezomed Plus sau modulul Piezomed Classic. Implantmed Plus II cu modulul Piezomed Plus II.
-
Modulul Piezomed: sunt piesele de mână ale modulelor SA-40 și SA-40 L compatibile cu dispozitivul Piezomed SA-320?
Nu, nu sunt compatibile.
-
Pot conecta Osstell Beacon la Implantmed Plus II?
Da, Osstell Beacon poate fi conectat prin intermediul unui adaptor USB-A/USB-C.
-
Pot conecta Implantmed Plus II la rețeaua Wi-Fi a cabinetului meu?
Da, dispozitivul acceptă orice rețea Wi-Fi criptată de 2,4 GHz.
-
Pot actualiza software-ul Implantmed Plus II mai târziu?
Da, pur și simplu creați un cont ioDent® și conectați dispozitivul la acesta. Vi se vor afișa apoi actualizările de software disponibile pentru dispozitivul dvs.
-
Funcția de scanare are capacitatea să detecteze implanturile pe care le folosesc?
Toate dispozitivele medicale cu un cod valid sunt afișate pe ecran după scanare. Când sunt conectate la ioDent®, materialele vor fi afișate în text simplu (nume, număr de serie etc.) în fișa pacientului.
-
Am nevoie de o conexiune la ioDent® pentru a utiliza Implantmed Plus II?
Nu, pentru utilizarea de bază a dispozitivului nu este necesar. Cu toate acestea, conectarea la ioDent® oferă funcții suplimentare. De exemplu, utilizatorii pot efectua singuri actualizări de software.
-
Ce tipuri de coduri pot fi scanate?
Coduri DataMatrix. Acestea conțin informații structurate despre produs, în conformitate cu Regulamentul privind dispozitivele medicale (MDR), și reprezintă un standard pentru dispozitivele medicale.
-
Se pot scana coduri QR și coduri de bare?
Nu. Codurile QR și codurile de bare nu pot fi scanate.
-
Pot integra sistemul SICAT Implant pe care îl dețin deja în ioDent®?
Da, integrarea cu SICAT Implant 2.0 este, în general, posibilă. Însă este necesară verificarea codului de licență.
-
Pot folosi software-ul meu pentru implanturi, Galileos (un pachet software SICAT Implant), pentru integrarea cu ioDent®?
Da, însă este necesară o actualizare.
-
Pot vedea pe portalul SICAT când este conectat ioDent®?
Da, pe portalul SICAT, la secțiunea: Profil > Setări cont > Conexiuni externe (Profile > Account Settings > External Connections)
-
De unde pot descărca SICAT Implant/SICAT Suite?
Puteți descărca produsul de aici https://www.sicat.com/wh, folosind codul promoțional inclus cu fiecare Implantmed.
-
Care sunt producătorii de implanturi ale căror protocoale sunt incluse în biblioteca de protocoale ioDent®?
În prezent, sunt acceptate următoarele protocoale: BIOHORIZONS IMPLANT SYSTEMS, INC; Bredent medical GmbH & Co. KG; CAMLOG Biotechnologies GmbH; Dentsply Implants Manufacturing GmbH; Implacil De Bortoli Material Odontológico S.A.; Nobel Biocare AB; Patent Medical AG; S.I.N. Implant System; Institut Straumann AG; Thommen Medical AG; TRI Dental Implants Int. AG
-
Ce se întâmplă dacă definesc un implant care poate fi planificat în SICAT Implant, dar nu există un protocol corespunzător în ioDent®?
În ioDent®, veți fi informat că protocolul nu este disponibil și că, pentru tratament, este setat și protocolul standard Implantmed.
-
Care sunt principalele avantaje ale programului SICAT Implant?
Pentru întrebări referitoare la SICAT în sine, vă invităm să luați legătura cu colegii de la SICAT, care vă vor oferi răspunsuri la întrebările specifice. Pe site-ul ecademy.sicat.com, sunt disponibile videoclipuri detaliate cu instrucțiuni de utilizare. În general, SICAT contribuie la optimizarea planificării și a tratamentului, prin ghidare digitală rapidă.
-
Ce tipuri de tuburi pot folosi cu dispozitivul meu?
-
Piezomed module Plus II (6)
-
Pot fi utilizate anse de alți producători în modulele Piezomed?
Nu
a. Filetul este diferit
b. Funcția de detectarea a anselor Piezomed nu poate detecta anse de altă marcă
-
Modulul Piezomed: se pot moderniza dispozitivele existente pe piață?
Da, dispozitivele pot fi modernizate de către utilizatori. Sunt posibile următoarele combinații: Implantmed Plus cu modulul Piezomed Plus sau modulul Piezomed Classic. Implantmed Plus II cu modulul Piezomed Plus II.
-
Modulul Piezomed: sunt piesele de mână ale modulelor SA-40 și SA-40 L compatibile cu dispozitivul Piezomed SA-320?
Nu, nu sunt compatibile.
-
Piezomed module: Compatibility of the handpieces with Piezomed module Plus II, Piezomed module Plus
and Piezomed Classic module.
Handpieces Piezomed module Plus II Piezomed module Plus Piezomed Classic module SA-40 yes yes yes SA-40 L (light version) yes yes no SA-320 no no no -
Se poate utiliza piesa SA-40 împreună cu modulul Piezomed Module Plus II?
Da, dar funcția de detectare automată a ansei trebuie dezactivată din submeniu. Setările, cum sunt grupa de putere, puterea și fluxul de răcire, trebuie configurate de utilizator conform fișei tehnice a ansei (inclusă cu fiecare ansă piezoelectrică W&H).
-
Cum conectez modulul Piezomed Module Plus II la Implantmed Plus II?
Pur și simplu, îndepărtați capacul lateral al echipamentului și glisați modulul, din spate, până când se fixează în poziție. Nu este necesar niciun cablu suplimentar.
-
Pot fi utilizate anse de alți producători în modulele Piezomed?
Turbine
-
Synea (10)
-
I've just bought a new W&H turbine? Should I call a serviceman for my unit to install it?
Yes, it is neccesary to properly adjust the air/water pressure and if applicable the voltage for the turbine.
-
At the moment I have a W&H turbine without light. How can I have light in my dental turbine?
If there is electricity at the end of turbine hose, you need a proper coupling and a turbine with light (example: RQ-24 and TK-97 L). If no electricity is available, the simplest way is to buy a W&H Alegra turbine with LED and its Roto Quick coupling (e.g.: TE-97 LQ and RQ-54). It doesn't require any other investment in the dental unit.
-
What is the maximum length of the rotary instruments?
25 mm for the standard turbine (XX-98 modells) and 21 mm for turbines with small head (XX-97 modells). When using longer rotary instruments the user must ensure by correct selection of the operating conditions, that there is no danger to the user, patient or third parties.
-
Is the new LED turbine from the Synea series e.g. compatible with the existing 924 couplings?
Yes, the new generation of turbines can be operated with both the existing and the new couplings.
-
My dental turbine doesn't have enough power/speed?
Please check drive air pressure on your unit and maintenance of the instruments (Synea with RQ connection = 3 +/- 0,3 bar, with Multiflex® connection 2,5 - 4 bar). For all Alegra turbines the pressure needs to be set between 2,5 - 2,8 bar.
-
Can my W&H turbine be processed via thermal disinfector?
Yes, providing that your W&H turbine carries the relevant thermal disinfection symbol and the thermo disinfector should have a drying cycle.
-
Can my W&H turbine be processed via sterilizer?
Yes, providing that your W&H turbine carries the relevant sterilization symbol.
-
Which burs can be used with W&H turbines?
Turbines with standard head sizes (models XX-98): 19 mm - 25 mm Turbines with small head sizes (models XX-97): 16 mm - 21 mm
The cutting part diameter for both head sizes is restricted to max. 2 mm. Please observe the technical data provided by the bur manufacturer. When using longer, rotary instruments for special treatment indications, the user must always select the operating conditions correctly in order to avoid placing the user, patient or anybody else at risk. -
How does the new W&H Click & Pull system work? How do I remove a dental turbine?
Use your thumb and index finger to retract the coupling's retention sleeve. The turbine can now be removed from the coupling without any problems. Pulling forcefully on the dental turbine can cause damage. If excessive force is required to remove the turbine, verify that the retention sleeve on the coupling can be moved without any problems.
-
Why is a 6-hole connection required for the use of an LED turbine?
A 6-hole connection for the turbine tubing is necessary to supply the turbine with electricity via electrical contacts. Other connections do not provide a power supply.
-
I've just bought a new W&H turbine? Should I call a serviceman for my unit to install it?
-
Primea (18)
-
In which versions can I purchase the Primea Advanced Air System? What are my installation options?
W&H can offer you the Primea Advanced Air System as an external add-on which you can connect to your dental unit. To do this, you need a spare supply hose for your dental unit, a minimum operating pressure of 5 bar and the possibility to connect the control system to the mains power supply. Please contact your authorized W&H service partner to check the inlet pressure of the additional air supply hose. The Primea Advanced Air Built-in System is integrated into a dental unit by selected dental unit manufacturers or their service technicians.
-
Can my Primea Advanced Air Turbine be prepared in a thermo washer disinfector?
Yes, if your Primea Advanced Air Turbine is labelled with the symbol "thermo washer disinfectable".
-
I have just purchased a new W&H Primea Advanced Air Turbine. How can I activate it?
It is not sufficient to purchase the Primea Advanced Air Turbine alone. In order to be able to operate it, you need the entire Primea Advanced Air System – consisting of a turbine, control system, user interface (UI) and the supply hose with Roto Quick coupling. You can add this as an add-on or built-in solution to/in your dental unit.
-
Are there different Primea Advanced Air Turbines?
Yes, the RG-97 L is equipped with an LED to illuminate the treatment area. The RK-97 L offers you shadowless illumination of the treatment site thanks to 5x Ring LED+. Both turbines have a mini-head and five spray ports, as well as a scratch-proof surface coating.
-
How can I integrate the Primea Advanced Air Built-in into my unit?
The Primea Advanced Air System is integrated into a dental unit by selected dental unit manufacturers or their service technicians.
-
How do I connect the Primea Advanced Air Add-on to my unit?
You can connect the external Primea Advanced Air System to the dental unit. To do this, you need a spare supply hose for your dental unit, a minimum operating pressure of 5 bar and the possibility to connect the control system to the mains power supply. Please contact an authorized W&H service partner to check the inlet pressure of the additional air supply hose.
-
Should I always call my authorized W&H service partner first before I send devices/instruments for repair?
If you are unsure of the cause of a problem, it is better to contact your authorized W&H service partner first and talk to a service technician. Together with the service technician, malfunctions may be resolved or it may be determined which items should be sent for repair.
-
What do I do if my Primea Advanced Air Turbine does not achieve sufficient power/speed?
Please contact your authorized W&H service partner to check the Primea Advanced Air System.
-
Which rotary instruments can be used with Primea Advanced Air Turbines?
Primea Advanced Air Turbines with small heads (Model RG-/RK--97 L): 16 mm to 21 mm The diameter of the working part is limited to a maximum of 2 mm. Follow the operating instructions of the manufacturer. When using longer or shorter rotary instruments the user must ensure by correct selection of the operating conditions, that there is no danger to the user, patient or third parties.
-
Which speed setting should I use on the Primea Advanced Air System?
Within the adjustable speed range of 60,000 to 320,000 revolutions, you can set the maximum permitted speed for the respective rotary instrument as specified by the manufacturer.
-
What is the maximum rotary instrument length approved by W&H for Primea Advanced Air Turbines?
Primea Advanced Air Turbines with small heads (Model RG-/RK-97 L): 16 mm to 21 mm
-
How does the Roto Quick coupling's W&H Click & Pull System work? How do I remove a Primea Advanced Air Turbine?
Pull the retention sleeve of the Roto Quick coupling back. The turbine can now be easily removed from the Roto Quick coupling. Forcefully pulling the Primea Advanced Air Turbine may cause damage. If excessive force is required to remove the turbine, make sure that the retention sleeve of the Roto Quick coupling can be moved without any problems.
-
How is the Primea Advanced Air System cleaned? How is it prepared?
For detailed instructions, see the chapter "Hygiene and maintenance" in the respective instructions for Use.
-
Does the Primea Advanced Air Built-in need to be serviced?
The built-in electronic components do not need to be serviced separately. However, regular servicing should be carried out together with the dental unit at the legally required intervals. This will ensure the proper functioning and safety of the components. For details, see the chapter "Service" in the respective instructions for Use.
-
Does the Primea Advanced Air Add-on need to be serviced?
Regular servicing of the Primea Advanced Air Add-on including the accessories should be carried out at least once every three years, unless shorter intervals are prescribed by law. The regular service must only be performed by an authorised W&H service partner. For details, see the chapter "Service" in the respective instructions for Use.
-
Can turbines from other manufacturers be operated in a Primea Advanced Air System?
This is possible. In order to operate standard turbines (from W&H and other manufacturers), the control system of the Primea Advanced Air System can be changed over to standard turbines mode. Turbines which can be connected to the standard 6-hole connection are then operated with a constant 3 bar operating pressure.
-
What features do the two operating modes "Power" and "Tactile" offer?
In "Power" mode, the set speed is also constantly maintained when pressure mounts on the rotary instrument, which translates to a higher removal rate. The "Tactile" mode permits a reduction in the removal rate when pressure mounts on the rotary instrument.
-
Poate fi sterilizată turbina mea Primea Advanced Air într-un sterilizator?
Da, dacă turbina dvs. Primea Advanced Air Turbine este etichetată cu simbolul "Sterilizabil până la temperatura indicată".
-
In which versions can I purchase the Primea Advanced Air System? What are my installation options?
-
Alegra (10)
-
I've just bought a new W&H turbine? Should I call a serviceman for my unit to install it?
Yes, it is neccesary to properly adjust the air/water pressure and if applicable the voltage for the turbine.
-
At the moment I have a W&H turbine without light. How can I have light in my dental turbine?
If there is electricity at the end of turbine hose, you need a proper coupling and a turbine with light (example: RQ-24 and TK-97 L). If no electricity is available, the simplest way is to buy a W&H Alegra turbine with LED and its Roto Quick coupling (e.g.: TE-97 LQ and RQ-54). It doesn't require any other investment in the dental unit.
-
What is the maximum length of the rotary instruments?
25 mm for the standard turbine (XX-98 modells) and 21 mm for turbines with small head (XX-97 modells). When using longer rotary instruments the user must ensure by correct selection of the operating conditions, that there is no danger to the user, patient or third parties.
-
My dental turbine doesn't have enough power/speed?
Please check drive air pressure on your unit and maintenance of the instruments (Synea with RQ connection = 3 +/- 0,3 bar, with Multiflex® connection 2,5 - 4 bar). For all Alegra turbines the pressure needs to be set between 2,5 - 2,8 bar.
-
Can my W&H turbine be processed via thermal disinfector?
Yes, providing that your W&H turbine carries the relevant thermal disinfection symbol and the thermo disinfector should have a drying cycle.
-
Can my W&H turbine be processed via sterilizer?
Yes, providing that your W&H turbine carries the relevant sterilization symbol.
-
Which burs can be used with W&H turbines?
Turbines with standard head sizes (models XX-98): 19 mm - 25 mm Turbines with small head sizes (models XX-97): 16 mm - 21 mm
The cutting part diameter for both head sizes is restricted to max. 2 mm. Please observe the technical data provided by the bur manufacturer. When using longer, rotary instruments for special treatment indications, the user must always select the operating conditions correctly in order to avoid placing the user, patient or anybody else at risk. -
How does the new W&H Click & Pull system work? How do I remove a dental turbine?
Use your thumb and index finger to retract the coupling's retention sleeve. The turbine can now be removed from the coupling without any problems. Pulling forcefully on the dental turbine can cause damage. If excessive force is required to remove the turbine, verify that the retention sleeve on the coupling can be moved without any problems.
-
Does the Alegra turbine handpiece also product light when used on optic dental units?
Alegra LED turbine handpieces are exclusively intended for use with Roto Quick couplings RQ-54 / RQ-53. Alegra LED turbine handpieces cannot be coupled onto Roto Quick couplings RQ-24 / RQ-34.
-
I have a new Alegra LED turbine handpiece, which is supplied with power by a Roto Quick coupling with integrated generator.
What should I pay attention to when servicing the Roto Quick coupling with generator?
The generator in the coupling is equipped with ball bearings, which must be lubricated once a month. This sufficiently removes any soiling from the generator.
-
I've just bought a new W&H turbine? Should I call a serviceman for my unit to install it?
Accesorii
-
SmartPeg (21)
-
De câte ori poate fi folosit SmartPeg?
În cazul SmartPeg-urilor reutilizabile, acestea pot fi folosite de până la 20 de cicluri de reprocesare, atâta timp cât integritatea și performanța acestora sunt menținute. Dacă în urma inspecției, se constată uzură, deformare, coroziune, filete deteriorate sau orice pierdere a compatibilității funcționale, SmartPeg-ul trebuie eliminat, chiar dacă a fost reprocesat de mai puțin de 20 de ori. SmartPeg-urile de unică folosință sunt destinate unei singure sesiuni de tratament și nu trebuie reprocesate. Translated with DeepL.com (free version)
-
Cât de des poate fi utilizat un SmartPeg?
Dispozitivul reutilizabil SmartPeg poate fi utilizat și reprocesat de până la 20 de ori, cu condiția ca integritatea și performanța acestuia să fie menținute. Dacă în urma inspecției se constată uzură, deformare, coroziune, fileturi deteriorate sau orice pierdere a compatibilității funcționale, dispozitivul SmartPeg trebuie eliminat, chiar dacă a fost reprocesat de mai puțin de 20 de ori.
-
Pot combina dispozitivele SmartPeg de unică folosință cu cele reutilizabile, în cadrul aceleiași ședințe de tratament?
Da. Puteți combina dispozitivele SmartPeg de unică folosință cu cele reutilizabile în cadrul aceleiași sesiuni de tratament. Singura cerință este să utilizați tipul corect de SmartPeg, corespunzător implantului sau bontului pe care îl măsurați. Faptul că dispozitivul SmartPeg este destinat utilizării unice sau reutilizării nu afectează măsurătoarea.
-
De ce SmartPeg-ul reutilizabil poate fi curățat și sterilizat, în timp ce SmartPeg-ul de unică folosință, nu, chiar dacă ambele sunt fabricate din aluminiu?
SmartPeg-ul reutilizabil este prevăzut cu un strat special de protecție la suprafață – unul pe corpul SmartPeg-ului și un alt strat protector negru pe magnet. Aceste straturi asigură rezistență la umiditate, agenți de curățare, uzură și coroziune, garantând că SmartPeg-ul își menține performanțele pe toată durata numărului validat de cicluri de reprocesare.
-
Un SmartPeg reutilizabil oferă aceleași valori ISQ ca un SmartPeg de unică folosință?
Da. Osstell este, de asemenea, singura companie care oferă, atât un SmartPeg steril de unică folosință, cât și unul reutilizabil, fabricat din aluminiu, având la bază peste 1.500 de studii publicate privind metoda ISQ. Utilizând scala ISQ, originală Osstell, și dispozitivele SmartPeg, nu sunt necesare tabele de conversie sau factori de corecție. Valoarea ISQ măsurată este valoarea reală, indiferent dacă se utilizează un SmartPeg de unică folosință sau unul reutilizabil.
-
Ce trebuie să fac dacă nu se afișează nicio valoare ISQ?
În cazul în care nu se afișează nicio valoare ISQ, aceasta înseamnă, de obicei, că vibrația dispozitivului SmartPeg nu este detectată clar de aparatul de măsurare. Această situație poate fi cauzată de mai mulți factori, dintre care majoritatea pot fi rezolvați cu ușurință:
1. Distanța față de aparatul de măsurare este prea mare
În mod normal, aparatul de măsurare trebuie ținut la o distanță de 2–4 mm față de magnetul situat în partea superioară a dispozitivului SmartPeg. Nu atingeți magnetul.
Dacă nu se afișează nicio valoare, apropiați dispozitivul de măsurare de magnet, până când se detectează un semnal.
2. Unghiul trebuie ajustat
Câmpul magnetic este direcțional, astfel încât orientarea dispozitivului de măsurare este importantă. Măsurătorile trebuie efectuate la un unghi de aproximativ 45 de grade față de partea superioară a dispozitivului SmartPeg.
3. Asigurați-vă că utilizați tipul corect de SmartPeg pentru implantul respectiv
Utilizați acest link pentru a găsi tipul corect de SmartPeg: Găsiți-mi SmartPeg-ul.
4. Atenuarea datorită țesuturilor moi
Țesuturile înconjurătoare pot, uneori, să atenueze vibrația mai mult decât se poate anticipa. Apropierea ușoară a dispozitivului de măsurare compensează, de obicei, acest efect.
5. Platforma implantului nu este curată Asigurați-vă că platforma implantului este curată și fără reziduuri, astfel încât SmartPeg să se așeze corect și să transmită vibrația.
6. SmartPeg este uzat sau deteriorat
• Dacă folosiți un SmartPeg reutilizabil, verificați-l înainte de utilizare.
Un SmartPeg reutilizabil care este uzat, îndoit sau are filetul deteriorat s-ar putea să nu vibreze corect – iar dispozitivul de măsurare s-ar putea să nu poată înregistra nicio valoare.
• Un SmartPeg de unică folosință este destinat utilizării într-o singură sesiune de tratament.
• Asigurați-vă că SmartPeg este așezat corect pe implant și că nu este în contact cu țesutul înconjurător, deoarece acest lucru poate împiedica detectarea precisă a semnalului și poate duce la lipsa valorii ISQ sau la o valoare incorectă a acesteia.
7. Interferențe electromagnetice
Sursele puternice de interferențe electromagnetice din apropiere pot perturba măsurarea. (Consultați secțiunea de întrebări frecvente privind „Zgomotul electromagnetic” pentru detalii.) -
Modul în care este strâns SmartPeg afectează valoarea ISQ?
Da. Valoarea ISQ poate fi afectată dacă SmartPeg-ul nu este montat corect. Pentru măsurători precise, SmartPeg-ul trebuie strâns manual cu ajutorul accesoriului SmartPeg Mount (aproximativ 4–6 Ncm). La acest cuplu de strângere, SmartPeg-ul are un contact ferm și stabil cu implantul, iar valoarea ISQ rezultată este constantă – de obicei în limita a ±2 unități ISQ. Dacă SmartPeg-ul nu este fixat corect:
• Contactul devine instabil, iar valoarea ISQ poate apărea fals scăzută.
Dacă se aplică o forță prea mare:
• Strângerea excesivă nu îmbunătățește precizia. Aceasta poate deteriora filetul SmartPeg.
- Un SmartPeg reutilizabil poate fi deteriorat, astfel încât să nu mai poată fi folosit din nou.
- Un SmartPeg de unică folosință sau reutilizabil poate suferi deformarea filetului, ceea ce duce la măsurători inexacte în timpul unei sesiuni sau, în cel mai rău caz, la ruperea SmartPeg-ului. -
Există vreo diferență vizuală între modelul SmartPeg de unică folosință și cel reutilizabil?
Da. SmartPeg-urile reutilizabile și cele de unică folosință sunt concepute să poată fi ușor diferențiate: • SmartPeg-ul de unică folosință este livrat în stare sterilă, are un magnet de culoarea aluminiului (fără contrast) și nu prezintă niciun marcaj de identificare realizat cu laser. • SmartPeg-ul reutilizabil are un magnet negru în partea superioară și poartă marcaje realizate cu laser, precum: tipul și numerele de lot, pentru identificare și trasabilitate.
-
Există studii care să demonstreze eficacitatea metodei Osstell?
În prezent, există mii de publicații științifice în care Osstell a fost utilizat ca parte a protocolului de studiu. Rezumatele acestora pot fi consultate prin intermediul bazei noastre de date, cu funcție de căutare, la adresa Osstell.com/scientific-database. În toate aceste studii s-au utilizat dispozitivele Osstell SmartPegs – fabricate din aluminiu –, care au asigurat standardul de măsurare consecvent și validat, pe care se bazează baza de dovezi ISQ.
-
Cum funcționează tehnologia Osstell?
Osstell măsoară stabilitatea implantului folosind analiza frecvenței de rezonanță (RFA). Micul dispozitiv SmartPeg, pe care îl fixați pe implant, este prevăzut cu un magnet minuscul în partea superioară. Când sonda emite impulsuri magnetice, SmartPeg începe să vibreze – similar cu o diapazon foarte mic.
Cum funcționează:
1. Sonda emite impulsuri magnetice.
Aceste impulsuri fac ca dispozitivul SmartPeg să vibreze fără contact fizic.
2. Intensitatea vibrației variază în funcție de stabilitatea implantului.
• Un implant stabil (suport osos bun) face ca dispozitivul SmartPeg să vibreze mai repede.
• Un implant mai puțin stabil îl face să vibreze mai lent.
3. Dispozitivul transformă această vibrație într-o valoare ISQ
Stabilitate mai mare → frecvență mai mare → valoare ISQ mai mare
Stabilitate mai mică → frecvență mai mică → valoare ISQ mai mică
Măsurarea durează mai puțin de o secundă, iar valoarea ISQ este afișată imediat. -
De ce este importantă scala Osstell ISQ?
Scala Osstell ISQ este relevantă întrucât oferă informații precise cu privire la evoluția osteointegrării, ajutând medicii să stabilească momentul optim pentru încărcarea protetică și asigurând un protocol chirurgical sigur.
-
De ce trebuie să efectuez două măsurători în același timp?
ISQ trebuie măsurat întotdeauna în două direcții, întrucât stabilitatea implantului poate varia în funcție de cantitatea de os înconjurător. Măsurarea din două unghiuri asigură captarea atât a direcției cu cea mai mare stabilitate, cât și a celei cu cea mai mică stabilitate. Pentru aceasta, efectuați o măsurătoare în direcția vestibular-linguală și una în direcția mezial-distală. Aceste două valori ISQ reprezintă cele două direcții de stabilitate. Uneori, valorile sunt identice; alteori, ele diferă. Dacă cele două măsurători diferă, luați întotdeauna în considerare valoarea ISQ cea mai mică, deoarece aceasta reflectă direcția cea mai slabă și, prin urmare, oferă cele mai relevante informații din punct de vedere clinic cu privire la stabilitatea implantului.
-
Cum se măsoară și se înregistrează ISQ la un bont protetic?
Valoarea ISQ la nivelul bontului se poate măsura și înregistra în OsstellConnect. Cu toate acestea, este esențial să se respecte procedura corectă de măsurare clinică pentru a obține valori precise.
Cum se măsoară ISQ la nivelul bontului:
1. Efectuați mai întâi o măsurătoare la nivelul implantului, înainte de a monta bontul.
2. Efectuați din nou măsurarea la nivelul bontului în cadrul aceleiași ședințe, utilizând un SmartPeg destinat măsurării la nivelul bontului.
3. Determinați decalajul notând diferența dintre ISQ-ul la nivelul implantului și cel la nivelul bontului. Acest decalaj rămâne stabil pe toată durata fazei de vindecare și trebuie luat în considerare la interpretarea măsurătorilor ulterioare, de follow-up.
4. La toate consultațiile ulterioare, efectuați măsurarea la nivelul bontului și aplicați același decalaj pentru a determina ISQ-ul real al implantului.
Cum se înregistrează măsurarea în OsstellConnect:
1. Accesați fișa pacientului, selectați pacientul și implantul din schema dentară. Faceți clic pe „Adăugați bont” (Add abutment), selectați bontul și salvați datele.
2. Selectați „Măsurați stabilitatea” (Measure stability)și alegeți „Bont” (Abutment) din secțiunea „Nivel de măsurare” ()Measurement level)). Efectuați măsurarea pe bont așa cum ați face la nivelul implantului și salvați.
3. Pentru a verifica evoluția stabilității, deschideți fila „Stabilitate” (Stability) și bifați opțiunea „Afișați bontul” (Show abutment). De asemenea, puteți activa opțiunea „Afișare detalii” (Show details) pentru a vizualiza valorile ISQ vestibular-lingual și mezial-distal. Pentru o explicație mai detaliată și îndrumări clinice, consultați articolul nostru de pe blog aici -
De ce obțin valori neașteptate sau absolut niciun rezultat de măsurare?
Există mai mulți factori care pot duce la valori ISQ neașteptate sau la imposibilitatea măsurării:
1. Utilizarea incorectă sau starea necorespunzătoare a dispozitivului SmartPeg
• SmartPeg-uri de unică folosință: Acestea sunt destinate unei singure sesiuni de tratament și unui singur pacient. Reutilizarea unui dispozitiv SmartPeg de unică folosință poate duce la uzura filetelor și la măsurători nesigure.
• SmartPeg-uri reutilizabile: Asigurați-vă că sunt curățate și inspectate corespunzător și că nu sunt deteriorate. Uzura, deformarea sau contaminarea pot afecta performanța.
2. Selectarea unui tip greșit de SmartPeg
Asigurați-vă că ați selectat SmartPeg-ul corect pentru sistemul de implant și conexiunea specifice. Consultați Ghidul de referință SmartPeg: osstell.com/smartpegguide 3. Prezența resturilor de țesut în zona de conexiune a implantului
Eventualele fragmente osoase, țesut moale sau lichide, prezente între interfața implantului și SmartPeg, pot împiedica poziționarea corectă a acestuia. Curățați temeinic zona de conectare protetică a implantului înainte de a atașa SmartPeg.
4. Interferențe electromagnetice Sursele de perturbații electromagnetice din apropiere pot perturba măsurarea. Dacă este posibil, îndepărtați sau opriți sursa de interferență.
5. Poziționarea incorectă a sondei Țineți vârful ansei la o distanță de 2–4 mm de SmartPeg și aliniați-l corect spre vârful acestuia. Nu atingeți magnetul. -
Utilizarea dispozitivului de măsurare Osstell afectează implantul?
Nu. Măsurarea cu Osstell nu afectează implantul. Dispozitivul de măsurare nu face decât să provoace vibrații la SmartPeg. Magnetul din SmartPeg este activat de impulsuri magnetice foarte slabe generate de dispozitiv. Tehnica este obiectivă, delicată și ușor replicabilă.
-
Cum selectez SmartPeg-ul potrivit pentru implantul meu?
Diferitele tipuri de implanturi necesită SmartPeg-uri specifice, pentru a asigura măsurarea exactă. Fiecare SmartPeg este adaptat unui anumit sistem de implant și unui anumit tip de conexiune. Puteți găsi SmartPeg-ul potrivit pentru implantul dumneavoastră în Ghidul de referință SmartPeg – faceți clic pe aici pentru a-l accesa.
-
Cum trebuie curățate și sterilizate SmartPeg-ul reutilizabil și accesoriul SmartPeg Mount?
Înainte de prima utilizare clinică și după fiecare utilizare, produsele reutilizabile trebuie curățate, dezinfectate și sterilizate conform instrucțiunilor din ghidul rapid disponibil aici: aici
-
Care dintre produsele Osstell pot fi sterilizate în autoclav?
Următoarele produse pot fi sterilizate (cu abur sub presiune de până la 135 °C):
– SmartPeg Mount
– SmartPeg-ul reutilizabil
– Sonda (IDx)
Vă rugăm să rețineți că produsele SmartPeg de unică folosință nu pot fi refolosite și nu pot fi sterilizate! -
Pot fi sterilizate SmartPeg-urile reutilizabile cu ciclului Prion?
Da, pot fi sterilizate cu ciclul „Prion”, la 134 °C timp de 18 minute.
-
„Am nevoie de o cheie dinamometrică pentru a strânge corect SmartPeg-ul?”
„Nu. Nu se recomandă utilizarea unei chei dinamometrice pentru strângerea dispozitivului SmartPeg. Dispozitivele SmartPeg trebuie montate întotdeauna cu ajutorul accesoriului SmartPeg Mount și strânse la un cuplu de aproximativ 4–6 Ncm, ceea ce se descrie de obicei ca fiind «strâns cu degetele». Astfel se asigură poziționarea corectă a dispozitivului SmartPeg, fără riscul de a deteriora conexiunea implantului.”
-
Cum se fixează SmartPeg-ul pe implant?
Indiferent că utilizați un SmartPeg de unică folosință sau unul reutilizabil, acesta trebuie montat întotdeauna cu ajutorul accesoriului SmartPeg Mount și strâns la un cuplu de aproximativ 4–6 Ncm, care este, în general, descris ca fiind „strâns cu degetele”. Astfel se asigură poziționarea corectă a SmartPeg-ului și precizia măsurătorilor ISQ, fără riscul de a deteriora conexiunea implantului sau SmartPeg-ul.
-
De câte ori poate fi folosit SmartPeg?
-
Instrumente Piezomed (6)
-
How long do the instruments last and when should they be replaced?
The useful service life of the instruments is determined by the length of time they are used reprocessing and sterilization cycles have only marginal influence.
a. Instruments need to be replaced
- If there is a loss of power while working
- Where there is visible damage on the working part or the shank of the instrument.
b. Saws must be replaced if
- Teeth are broken off
- Teeth are worn
c. Diamond-coated instruments must be replaced
- As soon as the diamond coating is worn
-
Pot fi utilizate anse de alți producători în modulele Piezomed?
Nu
a. Filetul este diferit
b. Funcția de detectarea a anselor Piezomed nu poate detecta anse de altă marcă
-
How are the instruments prepared for cleaning?
a. Preparation in the ultrasonic bath:
- Place the instruments in the instrument tray and lower into the ultrasonic bath.
- Use cleaning agents and disinfectants suitable for hand instruments (probes, mirrors, forceps, etc.).
DO NOT USE a bur bath, as this is very aggressive.
- After cleaning in the ultrasonic bath, rinse adequately with water to remove any remnants of the cleaning agent and disinfectant from the coolant channels.
- Blow dry with compressed air after rinsing
- Replace dried instruments in the instrument tray and package for sterilization
b. Mechanical preparation in the cleaning unit and disinfector
- Use the cleaning adapter for cleaning units and disinfectors (see Instructions for Use – Accessories)
- After preparation in the cleaning unit and disinfector, check that the coolant channels are dry and, if necessary, dry again with compressed air. -
Can Piezomed instruments also be used on third-party devices?
No
a. Different thread
b. Third-party device cannot find the correct resonance point – instrument fails to oscillate -
Is it necessary for the torque driver to engage audibly when tightening the instrument?
Yes, it must engage audibly once, as otherwise the required torque is not achieved.
-
Se poate utiliza piesa SA-40 împreună cu modulul Piezomed Module Plus II?
Da, dar funcția de detectare automată a ansei trebuie dezactivată din submeniu. Setările, cum sunt grupa de putere, puterea și fluxul de răcire, trebuie configurate de utilizator conform fișei tehnice a ansei (inclusă cu fiecare ansă piezoelectrică W&H).
-
How long do the instruments last and when should they be replaced?
Întreținere Piese
-
Assistina 301 plus (4)
-
How much air pressure does the Assistina 301 plus require in order to operate properly?
The pressure should be between 4 - 10 bar (58 to 145 psi).
-
Correct use of Assistina 301plus.
Please find a detailed description about first startup, regular use and maintenance in the user manual. Furthermore you can find video tutorials about Assistina 301plus.
-
How often should the filter in the supply air be changed (Assistina 301 plus)?
Annually or if the filter is full of debris (visual check - filter should be white). Please see video about changing the filter).
-
How can I check if my Assistina 301 is working correctly?
Check/replace the O-rings on the motor coupling/adapter (FAQ video Assistina O-rings). Ensure that the Start button is held depressed for at least 2 seconds when starting a programme. When the button is pressed, the green ball should be visible on the right in the functional display (cleaning solution). When the button is then released again, the green ball should be shown on the left. This procedure is also demonstrated in the FAQ video "Using Assistina". A detailed functional check can be found in the "weekly checklist" and in the "Test and functional check" section of the instructions for use.
-
How much air pressure does the Assistina 301 plus require in order to operate properly?
-
Assistina (1)
-
Correct use of Assistina 301plus.
Please find a detailed description about first startup, regular use and maintenance in the user manual. Furthermore you can find video tutorials about Assistina 301plus.
-
Correct use of Assistina 301plus.
-
Assistina ONE (7)
-
Care este diferența între lubrifierea cu ulei prin nebulizare și lubrifierea rotativă?
Rezultatul final este același, dar se obține într-un timp mai scurt. În sistemul de nebulizare, uleiul ajunge prima dată în adaptor, înainte de a fi pulverizat în interiorul piesei de mână. Se obține, astfel, o peliculă de ulei nebulizat pe toate suprafețele pieselor interne.
-
Cât de des trebuie înlocuit setul ONE Care?
Volumele de consum pentru fiecare instrument sunt de 0,07 ml de ulei și 0,07 ml de soluție de curățare. Aceasta oferă o capacitate maximă de 2 857 de cicluri. Valoarea se situează în jurul valorii de 2800 - 2850 de cicluri, deoarece este necesară și o umplere inițială, iar cantitatea de umplere a cartușelor este supusă anumitor toleranțe reglementate de standardele aplicabile.
-
Cum se înlocuiește setul ONE Care?
Cele două cartușe și filtrul pot fi schimbate cu ușurință, fără unelte. Fluidul este alimentat prin intermediul unor ace capilare care intră în partea de jos a cartușului, penetrează membrana cartușului și extrag astfel fluidul din cartuș. Cele două cartușe au forme diferite pentru a se asigura că cele două tipuri de fluid nu pot fi amestecate. În plus, instalarea incorectă a filtrului HEPA este posibilă doar prin forțare, ceea ce nu este necesar.
-
De ce nu există opțiunea de lubrifiere prin rotație la Assistina One?
Uleiul este pulverizat în Assistina One înainte ca acesta să pătrundă în piesă. Acest lucru asigură acoperirea completă a angrenajelor interne, eliminând necesitatea unei lubrifieri rotaționale, complexe din punct de vedere tehnic și consumatoare de timp. Cu toate acestea, nu există niciun impact negativ asupra calității procesului de întreținere.
-
Ce tipuri de adaptoare pentru piese sunt disponibile pentru Assistina One?
Toate adaptoarele care sunt potrivite pentru Assistina 3x2/3x3 pot fi utilizate și în Assistina One. Există o singură excepție: adaptorul RM/ISO pentru Assistina 3x2/3x3 nu este compatibil cu Assistina One. În schimb, pentru Assistina One pot fi utilizate adaptoarele Quick Connect ISO și RM. O prezentare generală a tuturor adaptoarelor este disponibilă pe site-ul web al produsului.
-
De ce are Assistina One filtru HEPA?
În timpul procesului de întreținere în Assistina One, piesele rotative sunt lubrifiate cu ulei, iar canalele de pulverizare sunt clătite cu soluție de curățare. Aceste lichide sunt introduse în piesele de mână cu ajutorul aerului comprimat, ceea ce produce aerosoli, care pot conține fluide corporale, potențial contaminate (de exemplu, sânge și salivă). Pentru a se asigura că aparatul nu prezintă niciun risc pentru utilizator, pacient sau terți, legislația prevede că aerosolii trebuie să fie împiedicați să se scurgă din aparat. Ne asigurăm de acest lucru prin utilizarea aspirației active și a filtrului HEPA. Acest filtru HEPA trebuie, bineînțeles, să fie schimbat în mod regulat, motiv pentru care durata de viață a filtrului este adaptată exact la capacitatea celor două cartușe. Prin urmare, filtrul HEPA ajunge la sfârșitul duratei sale de viață după ce a fost utilizat un set de cartușe.
-
Cum știe utilizatorul când trebuie să înlocuiască setul ONE Care?
Setul ONE Care trebuie înlocuit atunci când cartușele s-au epuizat. Nivelurile pot fi monitorizate cu ajutorul indicatoarelor de nivel de umplere ale aparatului, situate pe partea dreaptă și pe partea stângă.
-
Care este diferența între lubrifierea cu ulei prin nebulizare și lubrifierea rotativă?
Echipamente fără fir
-
Proxeo Twist Fără fir (13)
-
Is it possible to thermally disinfect and sterilize the Proxeo Twist Cordless?
The Proxeo Twist Cordless can be fully disinfected by wiping down. The handpiece sleeve can be sterilized and thermally disinfected. The foot control can also be disinfected by wiping down. You will find further information on this in the Instructions for use 50928/51037/51036.
-
Which prophy angle cups can be used with the Proxeo Twist Cordless?
The Proxeo Twist prophy angle cups from W&H fit the PL-40 H, providing outstanding benefits. You can, however, use all conventional prophy angle cups with the Doriot system.
For the PL-40 HW, different contra-angle heads are available. The PL-44 can be used with Proxeo Twist disposable Angle Cups from W&H or all commercially available disposable cups with Doriot system.
With the PL-64 contra-angle head you can use all Young (screw-in connection) cups.
The PL-66 W contra-angle head can be used with the LatchShort Prophy cups/brushes from W&H. -
How can I connect the Proxeo Twist Cordless with the foot control?
You will find further information on this in the Instructions for use 50928/51037.
-
Should I always call my service department first before I send devices/instruments for repair?
If you are unsure of the cause of the problem, it is better to contact your service department first and talk to a technician. Together with the technician, malfunctions can sometimes be resolved, or it may be determined which items need to be sent for repair.
-
How long is the battery life of the Proxeo Twist Cordless?
The rechargeable battery normally lasts for around one working day (8–12 patients with a polishing duration of 4–6 minutes [actual time the foot control is ‘on’]).
-
Is it the same rechargeable battery as a smartphone and what is the rechargeable battery’s capacity in comparison to the lifespan of the medical device? With a smartphone, I have to charge more often the older it gets.
The Proxeo Twist Cordless has a Li-ion rechargeable battery (like many smartphones). That said, you can’t compare the Proxeo Twist Cordless with a smartphone in this context. Smartphones have apps and updates installed on them which require more processing capacity and are also updated in the background. This shortens the life of the smartphone’s rechargeable battery and means it needs to be recharged more often, even though it only loses a small amount of capacity.
-
Are there any recommendations for charging the Proxeo Twist Cordless’ rechargeable battery?
Yes, we recommend charging the battery from time to time (e.g. during the lunch break), in order to achieve optimal performance.
-
How long does the foot control’s rechargeable battery last?
The rechargeable battery lasts approx. two months based on average use (5 working days with 8–12 patients/day for 4–6 minutes).
-
Can my W&H handpiece or contra-angle head be processed via thermal disinfector?
Yes, providing that your W&H handpiece or contra-angle head carries the thermal disinfection symbol.
-
Can I use the W&H Proxeo Twist Prophy Angle Cups and the LatchShort Prophy Cups and Brushes more than once?
No, both are single-use only. Please dispose the Prophy Angle Cups and the LatchShort Prophy Cups and Brushes after each treatment.
-
How can I disassemble the Young screw attachments from the PL-64 after the treatment?
For more information, please refer to Chapter 7 of the Instructions for Use 51037.
-
Can I use any Prophy cup that is available on the market for the PL-66 W?
In order to benefit from the advantages of the LatchShort system and to support the longevity of the contra-angle handpiece (PL-66 W), we recommend using W&H Prophy cups.
However, it is possible for Prophy cups from standard latch systems (with a shaft of 2.35 mm) to be used. -
Can my W&H handpiece or contra-angle head be processed via sterilizer?
Yes, providing that your W&H handpiece or contra-angle head carries the relevant sterilization symbol.
-
Is it possible to thermally disinfect and sterilize the Proxeo Twist Cordless?
Piese de mână drepte & contraunghi
-
Proxeo (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
-
Proxeo TWIST (5)
-
Can my W&H handpiece be processed via thermal disinfector and sterilizer?
Yes, providing that your W&H instrument carries the relevant thermal disinfection and sterilization symbol and the thermo disinfector should have a drying cycle.
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Why should I use a prophylaxis contra-angle handpiece (WP-66 W)?
Generally, prophylaxis contra-angle handpieces feature a seal on the head that prevents the ingress of saliva and paste. This extends the lifespan of the contra-angle handpiece.
The most effective removal of discolouration that at the same time protects the enamel and dentine as much as possible, depends on the polishing speed, i.e. revolutions per minute, among other things. With a speed ratio of 4:1 (four revolutions of the motor mean one revolution of the cup), the revolution speed of the Prophy cup is reduced. This means during application, there is no risk of paste splattering. -
Can I use any Prophy cup that is available on the market for the WP-66 W?
In order to benefit from the advantages of the LatchShort system and to support the longevity of the contra-angle handpiece (WP-66 W), we recommend using W&H Prophy cups. However, it is possible for Prophy cups from standard latch systems (with a shaft of 2.35 mm) to be used.
-
Should I call the service station before sending equipment for repair?
If you are not sure what the cause of the problem is, it is better to call the service station and talk to a technician. Together with the technician it can be determined what needs to be sent in and sometimes it is possible to even solve the problem right away.
-
Can my W&H handpiece be processed via thermal disinfector and sterilizer?
Tratarea Apei
-
Multidem (2)
-
Ce calitate a apei ar trebui să folosesc pentru sterilizatorul meu?
Calitatea apei pentru sterilizatoare este specificată la nivel internațional prin standarde și directive. Acest lucru se datorează faptului că sterilizatorul dvs. poate fi grav deteriorat de aceasta sau sterilizarea poate fi compromisă. Prin urmare, pentru sterilizatorul dumneavoastră W&H trebuie să utilizați numai apă cu o conductivitate mai mică de 10 μS/cm (în America de Nord) și de 15 μS/cm în restul lumii.Dispozitivele de tratare a apei Multidem și Osmo sunt special concepute pentru a asigura o apă adecvată (a se vedea capitolul "Calitatea apei" din instrucțiunile de utilizare).
-
Cartușul gol Multidem C27 poate fi returnat sau trebuie aruncat la gunoi de dumneavoastră (deșeuri menajere)?
Produsele trebuie eliminate în conformitate cu reglementările locale, regionale și naționale în vigoare; consultați recomandările privind eliminarea (secțiunea 11) cuprinse Fișa tehnică de siguranță a produsului.
-
Ce calitate a apei ar trebui să folosesc pentru sterilizatorul meu?
Detartraj Piezo
-
Proxeo Ultra (14)
-
Should I always call my service department first before I send devices/instruments for repair?
If you are unsure of the cause of the problem, it is better to contact your service department first and talk to a technician. Together with the technician, malfunctions can sometimes be resolved or it may be determined which items need to be sent for repair.
-
Are W&H tips (e.g. 1U) compatible with the new W&H high-speed mounting system?
No, they are not compatible because they have different threads.
-
How do I know which tip goes with which handpiece?
There are symbols on the tip, the tip changer and the handpiece identifying which system they belong to.
-
Are the new handpieces compatible with the PA-123/PA-115?
No, they are not compatible.
-
Are the new handpieces (PB-5 L, PB-5 L S, PB-5 L Q) compatible with the earlier generation of Piezo scalers?
No, they are not compatible.
-
Can the handpiece be prepared in the sterilizer?
Yes, as long as your handpiece or contra-angle handpiece carries the mark “sterilizable up to the stated temperature”.
-
Is the handpiece thermo washer disinfectable?
Yes, as long as your handpiece or contra-angle handpiece carries the mark “thermo washer disinfectable”.
-
Can I prepare the handpiece in universal hygiene devices (DAC-Universal)?
Yes.
-
How can I recharge the wireless foot control (PB-530)?
The wireless foot control can be recharged by plugging the power pack directly into the device.
-
How can I connect the wireless foot control with the Proxeo ULTRA PB-530?
Set the power regulator to “OFF”. Connect the control unit and the foot control using the cable. Press and hold the function button for 5 seconds.
-
Can I disinfect the tank by wiping it down?
Yes.
-
Can I disinfect the tank using the thermo washer?
No.
-
Can I treat patients with cardiac pacemakers with W&H Piezo scalers?
Yes, the compatibility of these devices has been assessed in application.
-
How long does the foot control’s rechargeable battery last?
The rechargeable battery lasts approx. two months based on average use (5 working days with 8–12 patients/day for 4–6 minutes).
-
Should I always call my service department first before I send devices/instruments for repair?
Sterilizatoare
-
MS Sterilizer (2)
-
Ce este validarea?
Validarea este o procedură de testare, efectuată de o autoritate oficială, care dovedește eficiența sterilizatorului dumneavoastră. Obligativitatea și frecvența validării sunt reglementate de normele locale/naționale în vigoare.
-
Trebuie să documentez rezultatele tuturor testelor periodice?
W&H recomandă consemnarea tuturor testelor, reparațiilor și modificărilor periodice ale sterilizatorului într-un Jurnal, inclusiv data și semnătura persoanei care a efectuat testul sau altă activitate. Înregistrarea obligatorie este reglementată de normele locale/naționale.
-
Ce este validarea?
-
Lisa Mini (12)
-
Trebuie să efectuez teste periodice pentru sterilizatoarele?
Sterilizatoarele dispun de cicluri de testare (testul B&D/Helix, testul Vacuum). Obligativitatea și frecvența testelor sunt reglementate de directivele locale/naționale.
-
Ar trebui să pun în pungă/împachetez instrumentele și cât timp pot să le păstrez după sterilizare?
Pentru a vă asigura că instrumentele dvs. sunt păstrate în stare sterilă de la sterilizarea lor până la utilizare, W&H recomandă să le împachetați, utilizând sisteme de bariere sterile adecvate (pungi/role de sterilizare) înainte de sterilizare. Perioada de depozitare recomandată poate fi găsită în recomandările producătorului materialului de împachetare. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Ce este validarea?
Validarea este o procedură de testare, efectuată de o autoritate oficială, care dovedește eficiența sterilizatorului dumneavoastră. Obligativitatea și frecvența validării sunt reglementate de normele locale/naționale în vigoare.
-
Ce calitate a apei ar trebui să folosesc pentru sterilizatorul meu?
Calitatea apei pentru sterilizatoare este specificată la nivel internațional prin standarde și directive. Acest lucru se datorează faptului că sterilizatorul dvs. poate fi grav deteriorat de aceasta sau sterilizarea poate fi compromisă. Prin urmare, pentru sterilizatorul dumneavoastră W&H trebuie să utilizați numai apă cu o conductivitate mai mică de 10 μS/cm (în America de Nord) și de 15 μS/cm în restul lumii.Dispozitivele de tratare a apei Multidem și Osmo sunt special concepute pentru a asigura o apă adecvată (a se vedea capitolul "Calitatea apei" din instrucțiunile de utilizare).
-
Pe cine ar trebui să contactez pentru validarea și mentenanța sterilizatorului meu?
Validarea se efectuează de către o autoritate oficială. Sarcinile de mentenanță trebuie efectuate de un tehnician calificat și autorizat.
-
Trebuie să documentez rezultatele tuturor testelor periodice?
W&H recomandă consemnarea tuturor testelor, reparațiilor și modificărilor periodice ale sterilizatorului într-un Jurnal, inclusiv data și semnătura persoanei care a efectuat testul sau altă activitate. Înregistrarea obligatorie este reglementată de normele locale/naționale.
-
Pentru cât timp este necesar să arhivez rapoartele testelor periodice?
Păstrați rezultatele în conformitate cu legile și ghidurile specifice țării/ regiunii dumneavoastră. Aceste registre pot fi utile și în eventualitatea intervențiilor de service.
-
Sterilizatorul meu are nevoie de vreo calificare/validare periodică?
Validarea cuprinde, în general, o serie de teste definite în standardele recunoscute la nivel internațional. La W&H, aceste teste încep, pentru fiecare echipament, cu testele din fabrică. În plus, documentația corespunzătoare este inclusă împreună cu echipamentul dumneavoastră. Înainte de punerea în funcțiune a dispozitivului la sediul dumneavoastră, tehnicianul efectuează verificări suplimentare pentru a se asigura că acesta funcționează corect. Apoi, în funcție de legile și directivele naționale/regionale, echipamentul trebuie să fie supus unei așa-numite evaluări reînnoite a performanțelor, la intervale regulate. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Cum se curăță rezervorul de apă și ce produse trebuie folosite?
Instrucțiunile de curățare a rezervorului de apă sunt furnizate în broșura Instrucțiuni de utilizare, capitolul „Întreținere”. În plus, aveți la dispoziție și o descriere video: - în sterilizator, în secțiunea Mentenanță din meniu.
-
Are nevoie Lisa Mini de o conexiune la aer comprimat?
Da, Lisa Mini necesită o conexiune la aer comprimat, fiind necesară pentru efectuarea ciclului de sterilizare. Specificațiile privind aerul comprimat sunt disponibile în Instrucțiunile de utilizare, capitolul „Date tehnice”.
-
Cum pot înlocui piesele consumabile ale sterilizatorului meu?
Instrucțiunile de înlocuire a consumabilelor sunt furnizate în Instrucțiunile de utilizare, capitolul „Întreținere”. În plus, aveți la dispoziție o descriere video: - în sterilizator, în secțiunea de întreținere din meniu.
-
Care este nivelul de zgomot maxim al sterilizatorului Lisa Mini?
Nivelul maxim de zgomot al Lisa Mini este de 64 dB.
-
Trebuie să efectuez teste periodice pentru sterilizatoarele?
-
Lara (24)
-
Există scurgeri de apă în partea din față a sterilizatorului.
CAUZĂ POSIBILĂ
Scurgeri prin garnitura ușii camerei.
SOLUȚIE
Curățați sau înlocuiți garnitura de etanșare a ușii. Curățați suprafața ușii dinspre cameră.
CAUZĂ POSIBILĂ
Scurgere internă.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Sterilizatorul rămâne oprit.
CAUZĂ POSIBILĂ
Întrerupătorul principal sau întrerupătorul de circuit al rețelei este oprit (OFF).
SOLUȚIE
Activați întrerupătorul principal sau întrerupătorul de rețea (ON).
CAUZĂ POSIBILĂ
Nu există tensiune la priză.
SOLUȚIE
Verificați alimentarea electrică a clădirii.
CAUZĂ POSIBILĂ
Cablul de alimentare nu este conectat corespunzător.
SOLUȚIE
Verificați și conectați corect cablul de alimentare. -
Trebuie să efectuez teste periodice pentru sterilizatoarele?
Sterilizatoarele dispun de cicluri de testare (testul B&D/Helix, testul Vacuum). Obligativitatea și frecvența testelor sunt reglementate de directivele locale/naționale.
-
Ar trebui să pun în pungă/împachetez instrumentele și cât timp pot să le păstrez după sterilizare?
Pentru a vă asigura că instrumentele dvs. sunt păstrate în stare sterilă de la sterilizarea lor până la utilizare, W&H recomandă să le împachetați, utilizând sisteme de bariere sterile adecvate (pungi/role de sterilizare) înainte de sterilizare. Perioada de depozitare recomandată poate fi găsită în recomandările producătorului materialului de împachetare. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Ce este validarea?
Validarea este o procedură de testare, efectuată de o autoritate oficială, care dovedește eficiența sterilizatorului dumneavoastră. Obligativitatea și frecvența validării sunt reglementate de normele locale/naționale în vigoare.
-
Ce calitate a apei ar trebui să folosesc pentru sterilizatorul meu?
Calitatea apei pentru sterilizatoare este specificată la nivel internațional prin standarde și directive. Acest lucru se datorează faptului că sterilizatorul dvs. poate fi grav deteriorat de aceasta sau sterilizarea poate fi compromisă. Prin urmare, pentru sterilizatorul dumneavoastră W&H trebuie să utilizați numai apă cu o conductivitate mai mică de 10 μS/cm (în America de Nord) și de 15 μS/cm în restul lumii.Dispozitivele de tratare a apei Multidem și Osmo sunt special concepute pentru a asigura o apă adecvată (a se vedea capitolul "Calitatea apei" din instrucțiunile de utilizare).
-
Cum se curăță rezervoarele de apă și ce produse ar trebui să folosesc?
Instrucțiunile de curățare a rezervorului de apă sunt furnizate în Instrucțiunile de utilizare, capitolul " Mentenanță". În plus, este disponibilă o descriere video: • în sterilizator, în secțiunea de mentenanță din meniu; • în secțiunea Consumabile și accesorii și secțiunea de Tutoriale Video a acestui website.
-
Cum pot schimba componentele consumabile ale sterilizatorului meu?
Instrucțiunile de înlocuire a consumabilelor sunt furnizate în Instrucțiunile de utilizare, capitolul " Mentenanță". În plus, este disponibilă o descriere video: • în sterilizator, în secțiunea de mentenanță din meniu; • în secțiunea Consumabile și accesorii și în secțiunea de Tutoriale Video a acestui website.
-
Pe cine ar trebui să contactez pentru validarea și mentenanța sterilizatorului meu?
Validarea se efectuează de către o autoritate oficială. Sarcinile de mentenanță trebuie efectuate de un tehnician calificat și autorizat.
-
Trebuie să documentez rezultatele tuturor testelor periodice?
W&H recomandă consemnarea tuturor testelor, reparațiilor și modificărilor periodice ale sterilizatorului într-un Jurnal, inclusiv data și semnătura persoanei care a efectuat testul sau altă activitate. Înregistrarea obligatorie este reglementată de normele locale/naționale.
-
Pentru cât timp este necesar să arhivez rapoartele testelor periodice?
Păstrați rezultatele în conformitate cu legile și ghidurile specifice țării/ regiunii dumneavoastră. Aceste registre pot fi utile și în eventualitatea intervențiilor de service.
-
În Meniu, la secțiunea Istoric cicluri, nu există nicio informație despre cicluri.
CAUZĂ POSIBILĂ
S-a înlocuit o placă electronică în timpul reparației.
SOLUȚIE
Memoria vechii plăci nu poate fi recuperată. Consultați istoricul ciclurilor de pe unitatea de memorie USB. -
Imprimanta pentru rapoartele de ciclu nu funcționează.
CAUZĂ POSIBILĂ
Imprimanta nu este conectată corect la sterilizator sau nu este conectată la rețeaua electrică.
SOLUȚIE
Verificați conexiunea de date și de alimentare a imprimantei. -
Sterilizatorul este conectat la un sistem automat de alimentare cu apă, nu există apă curată în rezervor, iar funcția de umplere automată a apei nu alimentează cu apă.
CAUZĂ POSIBILĂ
Sistemul de umplere cu apă nu este conectat.
SOLUȚIE
Conectați sistemul de umplere cu apă la sterilizator (a se vedea Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Când sistemul de umplere cu apă a încercat să umple rezervorul, apa a fost temporar indisponibilă.
SOLUȚIE
Deoarece umplerea rezervorului de apă este încercată doar o singură dată între două cicluri de execuție, acest eveniment inhibă alimentarea cu apă. Opriți sterilizatorul și apoi porniți-l din nou.
Verificați sistemul extern de alimentare cu apă.
Verificați dacă există scurgeri de apă din sterilizator.
CAUZĂ POSIBILĂ
Senzorul de nivel MIN defect în rezervorul de apă curată.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Avertizare privind mentenanța programată.
CAUZĂ POSIBILĂ
Trebuie efectuat programul de mentenanță.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Atenționare privind salvarea pe USB (fișiere HTML și SCL)
CAUZĂ POSIBILĂ
Dispozitivul de stocare USB este deconectat sau plin.
SOLUȚIE
Verificați prezența și starea stick-ului USB. Dacă problema persistă, contactați un partener de service autorizat W&H. -
Faza de sterilizare (PROCESUL) a unui ciclu de sterilizare a fost mai lungă decât se aștepta.
CAUZĂ POSIBILĂ
Temperatura camerei a scăzut sub pragul minim, iar software-ul a efectuat recuperarea cu succes.
SOLUȚIE
Așteptați finalizarea ciclului. Dacă problema apare frecvent, contactați un partener de service autorizat W&H. -
Sterilizatorul intră în "Modul Stand-by" imediat după pornire.
CAUZĂ POSIBILĂ
Unitatea a fost oprită în timp ce se afla în modul standby.
SOLUȚIE
Apăsați butonul Stand-by pentru a ieși. -
La pornirea unui ciclu, ușa camerei se blochează, dar se redeschide imediat. Apare mesajul "Open the door" (Deschideți ușa).
CAUZĂ POSIBILĂ
Garnitura de etanșare a ușii nu este pusă corect.
SOLUȚIE
Asigurați-vă că garnitura de etanșare a ușii este inserată corect pe întreaga circumferință.
CAUZĂ POSIBILĂ
Ușă blocată de obiecte externe sau de instrumentele încărcate.
SOLUȚIE
Îndepărtați orice obiect care se interpune cu ușa camerei. Verificați dacă există eventuale interferențe între ușă și elementele interioare ale camerei și/sau încărcătura. -
Pete, puncte sau modificări de culoare pe instrumente.
CAUZĂ POSIBILĂ
Instrumente clătite cu apă de la robinet și neuscate înainte de sterilizare. SOLUȚIE
Asigurați-vă că instrumentele sunt clătite cu apă distilată sau demineralizată și complet uscate înainte de a fi introduse în sterilizator.
CAUZĂ POSIBILĂ
Utilizarea apei de proastă calitate sau a apei care conține substanțe chimice.
SOLUȚIE
Goliți ambele rezervoare de apă. Folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare)..
CAUZĂ POSIBILĂ
Reziduuri organice sau chimice pe instrumente
SOLUȚIE
Curățați, clătiți și uscați instrumentele înainte de a le introduce în sterilizator (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare).
CAUZĂ POSIBILĂ
Mizerie în cameră, tăvi, suport de tăvi.
SOLUȚIE
Curățați camera și spălați elementele interioare ale camerei.
CAUZĂ POSIBILĂ
Contact între instrumente din diferite materiale.
SOLUȚIE
Asigurați-vă că instrumentele din materiale diferite nu se ating între ele (aluminiu, oțel carbon sau inox etc.); așezați-le pe tăvi sau casete diferite sau puneți-le în pungi (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Depuneri de calcar pe cameră.
SOLUTION
Curățați camera și folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare). -
La sfârșitul ciclului, există apă reziduală pe instrumente.
CAUZĂ POSIBILĂ
Sterilizatorul nu este poziționat corect la nivel. SOLUȚIE
Amplasați sterilizatorul pe o suprafață plană și nivelată.
CAUZĂ POSIBILĂ
Cameră supraîncărcată.
SOLUȚIE
Respectați limitele maxime de greutate a încărcăturii pentru fiecare tip de încărcătură. Folosiți întotdeauna suportul pentru tăvițe și casete.
CAUZĂ POSIBILĂ
Filtrul camerei de sterilizare este înfundat.
SOLUȚIE
Îndepărtați și curățați filtrul.
CAUZĂ POSIBILĂ
Capacul filtrului de cameră poziționat greșit.
SOLUȚIE
Montați corect capacul filtrului de cameră (consultați capitolul "Întreținerea lunară sau la 50 de cicluri - Curățarea filtrului de cameră" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Încărcătură amplasată incorect.
SOLUȚIE
Respectați recomandările enumerate în capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare. -
Sterilizatorul meu are nevoie de vreo calificare/validare periodică?
Validarea cuprinde, în general, o serie de teste definite în standardele recunoscute la nivel internațional. La W&H, aceste teste încep, pentru fiecare echipament, cu testele din fabrică. În plus, documentația corespunzătoare este inclusă împreună cu echipamentul dumneavoastră. Înainte de punerea în funcțiune a dispozitivului la sediul dumneavoastră, tehnicianul efectuează verificări suplimentare pentru a se asigura că acesta funcționează corect. Apoi, în funcție de legile și directivele naționale/regionale, echipamentul trebuie să fie supus unei așa-numite evaluări reînnoite a performanțelor, la intervale regulate. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Atenționare privind înlocuirea consumabilelor.
CAUZĂ POSIBILĂ
Este necesară înlocuirea unui consumabil.
SOLUȚIE
Comandați consumabilul solicitat (garnitura ușii, filtrul HEPA/bacteriologic) și înlocuiți-l. Consultați capitolul "Întreținere" din Instrucțiuni de utilizare. -
Care este nivelul maxim de zgomot al sterilizatorului Lara, Lara XL and Lexa Plus?
Nivelul maxim de zgomot al sterilizatoarelor Lara este 66,9 dB. Nivelul maxim de zgomot al sterilizatoarelor Lara XL/Lexa Plus este de 70 dB.
-
Există scurgeri de apă în partea din față a sterilizatorului.
-
Lina (24)
-
Există scurgeri de apă în partea din față a sterilizatorului.
CAUZĂ POSIBILĂ
Scurgeri prin garnitura ușii camerei.
SOLUȚIE
Curățați sau înlocuiți garnitura de etanșare a ușii. Curățați suprafața ușii dinspre cameră.
CAUZĂ POSIBILĂ
Scurgere internă.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Sterilizatorul rămâne oprit.
CAUZĂ POSIBILĂ
Întrerupătorul principal sau întrerupătorul de circuit al rețelei este oprit (OFF).
SOLUȚIE
Activați întrerupătorul principal sau întrerupătorul de rețea (ON).
CAUZĂ POSIBILĂ
Nu există tensiune la priză.
SOLUȚIE
Verificați alimentarea electrică a clădirii.
CAUZĂ POSIBILĂ
Cablul de alimentare nu este conectat corespunzător.
SOLUȚIE
Verificați și conectați corect cablul de alimentare. -
Trebuie să efectuez teste periodice pentru sterilizatoarele?
Sterilizatoarele dispun de cicluri de testare (testul B&D/Helix, testul Vacuum). Obligativitatea și frecvența testelor sunt reglementate de directivele locale/naționale.
-
Ar trebui să pun în pungă/împachetez instrumentele și cât timp pot să le păstrez după sterilizare?
Pentru a vă asigura că instrumentele dvs. sunt păstrate în stare sterilă de la sterilizarea lor până la utilizare, W&H recomandă să le împachetați, utilizând sisteme de bariere sterile adecvate (pungi/role de sterilizare) înainte de sterilizare. Perioada de depozitare recomandată poate fi găsită în recomandările producătorului materialului de împachetare. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Ce este validarea?
Validarea este o procedură de testare, efectuată de o autoritate oficială, care dovedește eficiența sterilizatorului dumneavoastră. Obligativitatea și frecvența validării sunt reglementate de normele locale/naționale în vigoare.
-
Ce calitate a apei ar trebui să folosesc pentru sterilizatorul meu?
Calitatea apei pentru sterilizatoare este specificată la nivel internațional prin standarde și directive. Acest lucru se datorează faptului că sterilizatorul dvs. poate fi grav deteriorat de aceasta sau sterilizarea poate fi compromisă. Prin urmare, pentru sterilizatorul dumneavoastră W&H trebuie să utilizați numai apă cu o conductivitate mai mică de 10 μS/cm (în America de Nord) și de 15 μS/cm în restul lumii.Dispozitivele de tratare a apei Multidem și Osmo sunt special concepute pentru a asigura o apă adecvată (a se vedea capitolul "Calitatea apei" din instrucțiunile de utilizare).
-
Cum se curăță rezervoarele de apă și ce produse ar trebui să folosesc?
Instrucțiunile de curățare a rezervorului de apă sunt furnizate în Instrucțiunile de utilizare, capitolul " Mentenanță". În plus, este disponibilă o descriere video: • în sterilizator, în secțiunea de mentenanță din meniu; • în secțiunea Consumabile și accesorii și secțiunea de Tutoriale Video a acestui website.
-
Cum pot schimba componentele consumabile ale sterilizatorului meu?
Instrucțiunile de înlocuire a consumabilelor sunt furnizate în Instrucțiunile de utilizare, capitolul " Mentenanță". În plus, este disponibilă o descriere video: • în sterilizator, în secțiunea de mentenanță din meniu; • în secțiunea Consumabile și accesorii și în secțiunea de Tutoriale Video a acestui website.
-
Pe cine ar trebui să contactez pentru validarea și mentenanța sterilizatorului meu?
Validarea se efectuează de către o autoritate oficială. Sarcinile de mentenanță trebuie efectuate de un tehnician calificat și autorizat.
-
Trebuie să documentez rezultatele tuturor testelor periodice?
W&H recomandă consemnarea tuturor testelor, reparațiilor și modificărilor periodice ale sterilizatorului într-un Jurnal, inclusiv data și semnătura persoanei care a efectuat testul sau altă activitate. Înregistrarea obligatorie este reglementată de normele locale/naționale.
-
Pentru cât timp este necesar să arhivez rapoartele testelor periodice?
Păstrați rezultatele în conformitate cu legile și ghidurile specifice țării/ regiunii dumneavoastră. Aceste registre pot fi utile și în eventualitatea intervențiilor de service.
-
În Meniu, la secțiunea Istoric cicluri, nu există nicio informație despre cicluri.
CAUZĂ POSIBILĂ
S-a înlocuit o placă electronică în timpul reparației.
SOLUȚIE
Memoria vechii plăci nu poate fi recuperată. Consultați istoricul ciclurilor de pe unitatea de memorie USB. -
Imprimanta pentru rapoartele de ciclu nu funcționează.
CAUZĂ POSIBILĂ
Imprimanta nu este conectată corect la sterilizator sau nu este conectată la rețeaua electrică.
SOLUȚIE
Verificați conexiunea de date și de alimentare a imprimantei. -
Sterilizatorul este conectat la un sistem automat de alimentare cu apă, nu există apă curată în rezervor, iar funcția de umplere automată a apei nu alimentează cu apă.
CAUZĂ POSIBILĂ
Sistemul de umplere cu apă nu este conectat.
SOLUȚIE
Conectați sistemul de umplere cu apă la sterilizator (a se vedea Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Când sistemul de umplere cu apă a încercat să umple rezervorul, apa a fost temporar indisponibilă.
SOLUȚIE
Deoarece umplerea rezervorului de apă este încercată doar o singură dată între două cicluri de execuție, acest eveniment inhibă alimentarea cu apă. Opriți sterilizatorul și apoi porniți-l din nou.
Verificați sistemul extern de alimentare cu apă.
Verificați dacă există scurgeri de apă din sterilizator.
CAUZĂ POSIBILĂ
Senzorul de nivel MIN defect în rezervorul de apă curată.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Avertizare privind mentenanța programată.
CAUZĂ POSIBILĂ
Trebuie efectuat programul de mentenanță.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Atenționare privind salvarea pe USB (fișiere HTML și SCL)
CAUZĂ POSIBILĂ
Dispozitivul de stocare USB este deconectat sau plin.
SOLUȚIE
Verificați prezența și starea stick-ului USB. Dacă problema persistă, contactați un partener de service autorizat W&H. -
Faza de sterilizare (PROCESUL) a unui ciclu de sterilizare a fost mai lungă decât se aștepta.
CAUZĂ POSIBILĂ
Temperatura camerei a scăzut sub pragul minim, iar software-ul a efectuat recuperarea cu succes.
SOLUȚIE
Așteptați finalizarea ciclului. Dacă problema apare frecvent, contactați un partener de service autorizat W&H. -
Sterilizatorul intră în "Modul Stand-by" imediat după pornire.
CAUZĂ POSIBILĂ
Unitatea a fost oprită în timp ce se afla în modul standby.
SOLUȚIE
Apăsați butonul Stand-by pentru a ieși. -
La pornirea unui ciclu, ușa camerei se blochează, dar se redeschide imediat. Apare mesajul "Open the door" (Deschideți ușa).
CAUZĂ POSIBILĂ
Garnitura de etanșare a ușii nu este pusă corect.
SOLUȚIE
Asigurați-vă că garnitura de etanșare a ușii este inserată corect pe întreaga circumferință.
CAUZĂ POSIBILĂ
Ușă blocată de obiecte externe sau de instrumentele încărcate.
SOLUȚIE
Îndepărtați orice obiect care se interpune cu ușa camerei. Verificați dacă există eventuale interferențe între ușă și elementele interioare ale camerei și/sau încărcătura. -
Pete, puncte sau modificări de culoare pe instrumente.
CAUZĂ POSIBILĂ
Instrumente clătite cu apă de la robinet și neuscate înainte de sterilizare. SOLUȚIE
Asigurați-vă că instrumentele sunt clătite cu apă distilată sau demineralizată și complet uscate înainte de a fi introduse în sterilizator.
CAUZĂ POSIBILĂ
Utilizarea apei de proastă calitate sau a apei care conține substanțe chimice.
SOLUȚIE
Goliți ambele rezervoare de apă. Folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare)..
CAUZĂ POSIBILĂ
Reziduuri organice sau chimice pe instrumente
SOLUȚIE
Curățați, clătiți și uscați instrumentele înainte de a le introduce în sterilizator (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare).
CAUZĂ POSIBILĂ
Mizerie în cameră, tăvi, suport de tăvi.
SOLUȚIE
Curățați camera și spălați elementele interioare ale camerei.
CAUZĂ POSIBILĂ
Contact între instrumente din diferite materiale.
SOLUȚIE
Asigurați-vă că instrumentele din materiale diferite nu se ating între ele (aluminiu, oțel carbon sau inox etc.); așezați-le pe tăvi sau casete diferite sau puneți-le în pungi (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Depuneri de calcar pe cameră.
SOLUTION
Curățați camera și folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare). -
La sfârșitul ciclului, există apă reziduală pe instrumente.
CAUZĂ POSIBILĂ
Sterilizatorul nu este poziționat corect la nivel. SOLUȚIE
Amplasați sterilizatorul pe o suprafață plană și nivelată.
CAUZĂ POSIBILĂ
Cameră supraîncărcată.
SOLUȚIE
Respectați limitele maxime de greutate a încărcăturii pentru fiecare tip de încărcătură. Folosiți întotdeauna suportul pentru tăvițe și casete.
CAUZĂ POSIBILĂ
Filtrul camerei de sterilizare este înfundat.
SOLUȚIE
Îndepărtați și curățați filtrul.
CAUZĂ POSIBILĂ
Capacul filtrului de cameră poziționat greșit.
SOLUȚIE
Montați corect capacul filtrului de cameră (consultați capitolul "Întreținerea lunară sau la 50 de cicluri - Curățarea filtrului de cameră" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Încărcătură amplasată incorect.
SOLUȚIE
Respectați recomandările enumerate în capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare. -
Sterilizatorul meu are nevoie de vreo calificare/validare periodică?
Validarea cuprinde, în general, o serie de teste definite în standardele recunoscute la nivel internațional. La W&H, aceste teste încep, pentru fiecare echipament, cu testele din fabrică. În plus, documentația corespunzătoare este inclusă împreună cu echipamentul dumneavoastră. Înainte de punerea în funcțiune a dispozitivului la sediul dumneavoastră, tehnicianul efectuează verificări suplimentare pentru a se asigura că acesta funcționează corect. Apoi, în funcție de legile și directivele naționale/regionale, echipamentul trebuie să fie supus unei așa-numite evaluări reînnoite a performanțelor, la intervale regulate. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Atenționare privind înlocuirea consumabilelor.
CAUZĂ POSIBILĂ
Este necesară înlocuirea unui consumabil.
SOLUȚIE
Comandați consumabilul solicitat (garnitura ușii, filtrul HEPA/bacteriologic) și înlocuiți-l. Consultați capitolul "Întreținere" din Instrucțiuni de utilizare. -
Which is the maximum noise level of Lina?
The max. noise level of Lina is 65.5 dB.
-
Există scurgeri de apă în partea din față a sterilizatorului.
-
Lara XXL (25)
-
Există scurgeri de apă în partea din față a sterilizatorului.
CAUZĂ POSIBILĂ
Scurgeri prin garnitura ușii camerei.
SOLUȚIE
Curățați sau înlocuiți garnitura de etanșare a ușii. Curățați suprafața ușii dinspre cameră.
CAUZĂ POSIBILĂ
Scurgere internă.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Sterilizatorul rămâne oprit.
CAUZĂ POSIBILĂ
Întrerupătorul principal sau întrerupătorul de circuit al rețelei este oprit (OFF).
SOLUȚIE
Activați întrerupătorul principal sau întrerupătorul de rețea (ON).
CAUZĂ POSIBILĂ
Nu există tensiune la priză.
SOLUȚIE
Verificați alimentarea electrică a clădirii.
CAUZĂ POSIBILĂ
Cablul de alimentare nu este conectat corespunzător.
SOLUȚIE
Verificați și conectați corect cablul de alimentare. -
Trebuie să efectuez teste periodice pentru sterilizatoarele?
Sterilizatoarele dispun de cicluri de testare (testul B&D/Helix, testul Vacuum). Obligativitatea și frecvența testelor sunt reglementate de directivele locale/naționale.
-
Ar trebui să pun în pungă/împachetez instrumentele și cât timp pot să le păstrez după sterilizare?
Pentru a vă asigura că instrumentele dvs. sunt păstrate în stare sterilă de la sterilizarea lor până la utilizare, W&H recomandă să le împachetați, utilizând sisteme de bariere sterile adecvate (pungi/role de sterilizare) înainte de sterilizare. Perioada de depozitare recomandată poate fi găsită în recomandările producătorului materialului de împachetare. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Ce este validarea?
Validarea este o procedură de testare, efectuată de o autoritate oficială, care dovedește eficiența sterilizatorului dumneavoastră. Obligativitatea și frecvența validării sunt reglementate de normele locale/naționale în vigoare.
-
Ce calitate a apei ar trebui să folosesc pentru sterilizatorul meu?
Calitatea apei pentru sterilizatoare este specificată la nivel internațional prin standarde și directive. Acest lucru se datorează faptului că sterilizatorul dvs. poate fi grav deteriorat de aceasta sau sterilizarea poate fi compromisă. Prin urmare, pentru sterilizatorul dumneavoastră W&H trebuie să utilizați numai apă cu o conductivitate mai mică de 10 μS/cm (în America de Nord) și de 15 μS/cm în restul lumii.Dispozitivele de tratare a apei Multidem și Osmo sunt special concepute pentru a asigura o apă adecvată (a se vedea capitolul "Calitatea apei" din instrucțiunile de utilizare).
-
Cum se curăță rezervoarele de apă și ce produse ar trebui să folosesc?
Instrucțiunile de curățare a rezervorului de apă sunt furnizate în Instrucțiunile de utilizare, capitolul " Mentenanță". În plus, este disponibilă o descriere video: • în sterilizator, în secțiunea de mentenanță din meniu; • în secțiunea Consumabile și accesorii și secțiunea de Tutoriale Video a acestui website.
-
Pe cine ar trebui să contactez pentru validarea și mentenanța sterilizatorului meu?
Validarea se efectuează de către o autoritate oficială. Sarcinile de mentenanță trebuie efectuate de un tehnician calificat și autorizat.
-
Trebuie să documentez rezultatele tuturor testelor periodice?
W&H recomandă consemnarea tuturor testelor, reparațiilor și modificărilor periodice ale sterilizatorului într-un Jurnal, inclusiv data și semnătura persoanei care a efectuat testul sau altă activitate. Înregistrarea obligatorie este reglementată de normele locale/naționale.
-
Pentru cât timp este necesar să arhivez rapoartele testelor periodice?
Păstrați rezultatele în conformitate cu legile și ghidurile specifice țării/ regiunii dumneavoastră. Aceste registre pot fi utile și în eventualitatea intervențiilor de service.
-
În Meniu, la secțiunea Istoric cicluri, nu există nicio informație despre cicluri.
CAUZĂ POSIBILĂ
S-a înlocuit o placă electronică în timpul reparației.
SOLUȚIE
Memoria vechii plăci nu poate fi recuperată. Consultați istoricul ciclurilor de pe unitatea de memorie USB. -
Imprimanta pentru rapoartele de ciclu nu funcționează.
CAUZĂ POSIBILĂ
Imprimanta nu este conectată corect la sterilizator sau nu este conectată la rețeaua electrică.
SOLUȚIE
Verificați conexiunea de date și de alimentare a imprimantei. -
Sterilizatorul este conectat la un sistem automat de alimentare cu apă, nu există apă curată în rezervor, iar funcția de umplere automată a apei nu alimentează cu apă.
CAUZĂ POSIBILĂ
Sistemul de umplere cu apă nu este conectat.
SOLUȚIE
Conectați sistemul de umplere cu apă la sterilizator (a se vedea Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Când sistemul de umplere cu apă a încercat să umple rezervorul, apa a fost temporar indisponibilă.
SOLUȚIE
Deoarece umplerea rezervorului de apă este încercată doar o singură dată între două cicluri de execuție, acest eveniment inhibă alimentarea cu apă. Opriți sterilizatorul și apoi porniți-l din nou.
Verificați sistemul extern de alimentare cu apă.
Verificați dacă există scurgeri de apă din sterilizator.
CAUZĂ POSIBILĂ
Senzorul de nivel MIN defect în rezervorul de apă curată.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Avertizare privind mentenanța programată.
CAUZĂ POSIBILĂ
Trebuie efectuat programul de mentenanță.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Atenționare privind salvarea pe USB (fișiere HTML și SCL)
CAUZĂ POSIBILĂ
Dispozitivul de stocare USB este deconectat sau plin.
SOLUȚIE
Verificați prezența și starea stick-ului USB. Dacă problema persistă, contactați un partener de service autorizat W&H. -
Faza de sterilizare (PROCESUL) a unui ciclu de sterilizare a fost mai lungă decât se aștepta.
CAUZĂ POSIBILĂ
Temperatura camerei a scăzut sub pragul minim, iar software-ul a efectuat recuperarea cu succes.
SOLUȚIE
Așteptați finalizarea ciclului. Dacă problema apare frecvent, contactați un partener de service autorizat W&H. -
Sterilizatorul intră în "Modul Stand-by" imediat după pornire.
CAUZĂ POSIBILĂ
Unitatea a fost oprită în timp ce se afla în modul standby.
SOLUȚIE
Apăsați butonul Stand-by pentru a ieși. -
La pornirea unui ciclu, ușa camerei se blochează, dar se redeschide imediat. Apare mesajul "Open the door" (Deschideți ușa).
CAUZĂ POSIBILĂ
Garnitura de etanșare a ușii nu este pusă corect.
SOLUȚIE
Asigurați-vă că garnitura de etanșare a ușii este inserată corect pe întreaga circumferință.
CAUZĂ POSIBILĂ
Ușă blocată de obiecte externe sau de instrumentele încărcate.
SOLUȚIE
Îndepărtați orice obiect care se interpune cu ușa camerei. Verificați dacă există eventuale interferențe între ușă și elementele interioare ale camerei și/sau încărcătura. -
Pete, puncte sau modificări de culoare pe instrumente.
CAUZĂ POSIBILĂ
Instrumente clătite cu apă de la robinet și neuscate înainte de sterilizare. SOLUȚIE
Asigurați-vă că instrumentele sunt clătite cu apă distilată sau demineralizată și complet uscate înainte de a fi introduse în sterilizator.
CAUZĂ POSIBILĂ
Utilizarea apei de proastă calitate sau a apei care conține substanțe chimice.
SOLUȚIE
Goliți ambele rezervoare de apă. Folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare)..
CAUZĂ POSIBILĂ
Reziduuri organice sau chimice pe instrumente
SOLUȚIE
Curățați, clătiți și uscați instrumentele înainte de a le introduce în sterilizator (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare).
CAUZĂ POSIBILĂ
Mizerie în cameră, tăvi, suport de tăvi.
SOLUȚIE
Curățați camera și spălați elementele interioare ale camerei.
CAUZĂ POSIBILĂ
Contact între instrumente din diferite materiale.
SOLUȚIE
Asigurați-vă că instrumentele din materiale diferite nu se ating între ele (aluminiu, oțel carbon sau inox etc.); așezați-le pe tăvi sau casete diferite sau puneți-le în pungi (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Depuneri de calcar pe cameră.
SOLUTION
Curățați camera și folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare). -
La sfârșitul ciclului, există apă reziduală pe instrumente.
CAUZĂ POSIBILĂ
Sterilizatorul nu este poziționat corect la nivel. SOLUȚIE
Amplasați sterilizatorul pe o suprafață plană și nivelată.
CAUZĂ POSIBILĂ
Cameră supraîncărcată.
SOLUȚIE
Respectați limitele maxime de greutate a încărcăturii pentru fiecare tip de încărcătură. Folosiți întotdeauna suportul pentru tăvițe și casete.
CAUZĂ POSIBILĂ
Filtrul camerei de sterilizare este înfundat.
SOLUȚIE
Îndepărtați și curățați filtrul.
CAUZĂ POSIBILĂ
Capacul filtrului de cameră poziționat greșit.
SOLUȚIE
Montați corect capacul filtrului de cameră (consultați capitolul "Întreținerea lunară sau la 50 de cicluri - Curățarea filtrului de cameră" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Încărcătură amplasată incorect.
SOLUȚIE
Respectați recomandările enumerate în capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare. -
Sterilizatorul meu are nevoie de vreo calificare/validare periodică?
Validarea cuprinde, în general, o serie de teste definite în standardele recunoscute la nivel internațional. La W&H, aceste teste încep, pentru fiecare echipament, cu testele din fabrică. În plus, documentația corespunzătoare este inclusă împreună cu echipamentul dumneavoastră. Înainte de punerea în funcțiune a dispozitivului la sediul dumneavoastră, tehnicianul efectuează verificări suplimentare pentru a se asigura că acesta funcționează corect. Apoi, în funcție de legile și directivele naționale/regionale, echipamentul trebuie să fie supus unei așa-numite evaluări reînnoite a performanțelor, la intervale regulate. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Atenționare privind înlocuirea consumabilelor.
CAUZĂ POSIBILĂ
Este necesară înlocuirea unui consumabil.
SOLUȚIE
Comandați consumabilul solicitat (garnitura ușii, filtrul HEPA/bacteriologic) și înlocuiți-l. Consultați capitolul "Întreținere" din Instrucțiuni de utilizare. -
Cum pot înlocui consumabilele sterilizatorului meu?
Instrucțiunile de înlocuire a consumabilelor sunt furnizate în Instrucțiunile de utilizare, capitolul „Întreținere”. În plus, veți găsi o descriere video în meniul sterilizatorui, la secțiunea de mentenanță, dar și în secțiunea Consumabile și Accesorii și secțiunea Video Tutorial a acestui site.
-
Care este nivelul de zgomot maxim al sterilizatorului Lara XXL?
Nivelul maxim al zgomotului produs de Lara XXL este de 70 dB.
-
La sfârșitul ciclului, se observă apă reziduală pe dispozitivele încărcate.
CAUZA POSIBILĂ Sterilizatorul nu este poziționat corespunzător, pe plan orizontal. SOLUȚIE Repoziționați sterilizatorul pe o suprafață plană, verificată cu nivela. CAUZA POSIBILĂ Cameră supraîncărcată. SOLUȚIE Respectați limitele de greutate maximă indicate fiecărui tip de încărcătură. Utilizați întotdeauna suportul pentru tăvi și casete potrivit. CAUZA POSIBILĂ Filtrele camerei de sterilizare sunt înfundate. SOLUȚIE Scoateți și curățați filtrele camerei. CAUZA POSIBILĂ Încărcătură plasată incorect. SOLUȚIE Urmați recomandările enumerate în capitolul „Întreținerea și pregătirea încărcăturii” din Instrucțiunile de utilizare.
-
Există scurgeri de apă în partea din față a sterilizatorului.
-
Lara XL (24)
-
Există scurgeri de apă în partea din față a sterilizatorului.
CAUZĂ POSIBILĂ
Scurgeri prin garnitura ușii camerei.
SOLUȚIE
Curățați sau înlocuiți garnitura de etanșare a ușii. Curățați suprafața ușii dinspre cameră.
CAUZĂ POSIBILĂ
Scurgere internă.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Sterilizatorul rămâne oprit.
CAUZĂ POSIBILĂ
Întrerupătorul principal sau întrerupătorul de circuit al rețelei este oprit (OFF).
SOLUȚIE
Activați întrerupătorul principal sau întrerupătorul de rețea (ON).
CAUZĂ POSIBILĂ
Nu există tensiune la priză.
SOLUȚIE
Verificați alimentarea electrică a clădirii.
CAUZĂ POSIBILĂ
Cablul de alimentare nu este conectat corespunzător.
SOLUȚIE
Verificați și conectați corect cablul de alimentare. -
Trebuie să efectuez teste periodice pentru sterilizatoarele?
Sterilizatoarele dispun de cicluri de testare (testul B&D/Helix, testul Vacuum). Obligativitatea și frecvența testelor sunt reglementate de directivele locale/naționale.
-
Ar trebui să pun în pungă/împachetez instrumentele și cât timp pot să le păstrez după sterilizare?
Pentru a vă asigura că instrumentele dvs. sunt păstrate în stare sterilă de la sterilizarea lor până la utilizare, W&H recomandă să le împachetați, utilizând sisteme de bariere sterile adecvate (pungi/role de sterilizare) înainte de sterilizare. Perioada de depozitare recomandată poate fi găsită în recomandările producătorului materialului de împachetare. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Ce este validarea?
Validarea este o procedură de testare, efectuată de o autoritate oficială, care dovedește eficiența sterilizatorului dumneavoastră. Obligativitatea și frecvența validării sunt reglementate de normele locale/naționale în vigoare.
-
Ce calitate a apei ar trebui să folosesc pentru sterilizatorul meu?
Calitatea apei pentru sterilizatoare este specificată la nivel internațional prin standarde și directive. Acest lucru se datorează faptului că sterilizatorul dvs. poate fi grav deteriorat de aceasta sau sterilizarea poate fi compromisă. Prin urmare, pentru sterilizatorul dumneavoastră W&H trebuie să utilizați numai apă cu o conductivitate mai mică de 10 μS/cm (în America de Nord) și de 15 μS/cm în restul lumii.Dispozitivele de tratare a apei Multidem și Osmo sunt special concepute pentru a asigura o apă adecvată (a se vedea capitolul "Calitatea apei" din instrucțiunile de utilizare).
-
Cum se curăță rezervoarele de apă și ce produse ar trebui să folosesc?
Instrucțiunile de curățare a rezervorului de apă sunt furnizate în Instrucțiunile de utilizare, capitolul " Mentenanță". În plus, este disponibilă o descriere video: • în sterilizator, în secțiunea de mentenanță din meniu; • în secțiunea Consumabile și accesorii și secțiunea de Tutoriale Video a acestui website.
-
Cum pot schimba componentele consumabile ale sterilizatorului meu?
Instrucțiunile de înlocuire a consumabilelor sunt furnizate în Instrucțiunile de utilizare, capitolul " Mentenanță". În plus, este disponibilă o descriere video: • în sterilizator, în secțiunea de mentenanță din meniu; • în secțiunea Consumabile și accesorii și în secțiunea de Tutoriale Video a acestui website.
-
Pe cine ar trebui să contactez pentru validarea și mentenanța sterilizatorului meu?
Validarea se efectuează de către o autoritate oficială. Sarcinile de mentenanță trebuie efectuate de un tehnician calificat și autorizat.
-
Trebuie să documentez rezultatele tuturor testelor periodice?
W&H recomandă consemnarea tuturor testelor, reparațiilor și modificărilor periodice ale sterilizatorului într-un Jurnal, inclusiv data și semnătura persoanei care a efectuat testul sau altă activitate. Înregistrarea obligatorie este reglementată de normele locale/naționale.
-
Pentru cât timp este necesar să arhivez rapoartele testelor periodice?
Păstrați rezultatele în conformitate cu legile și ghidurile specifice țării/ regiunii dumneavoastră. Aceste registre pot fi utile și în eventualitatea intervențiilor de service.
-
În Meniu, la secțiunea Istoric cicluri, nu există nicio informație despre cicluri.
CAUZĂ POSIBILĂ
S-a înlocuit o placă electronică în timpul reparației.
SOLUȚIE
Memoria vechii plăci nu poate fi recuperată. Consultați istoricul ciclurilor de pe unitatea de memorie USB. -
Imprimanta pentru rapoartele de ciclu nu funcționează.
CAUZĂ POSIBILĂ
Imprimanta nu este conectată corect la sterilizator sau nu este conectată la rețeaua electrică.
SOLUȚIE
Verificați conexiunea de date și de alimentare a imprimantei. -
Sterilizatorul este conectat la un sistem automat de alimentare cu apă, nu există apă curată în rezervor, iar funcția de umplere automată a apei nu alimentează cu apă.
CAUZĂ POSIBILĂ
Sistemul de umplere cu apă nu este conectat.
SOLUȚIE
Conectați sistemul de umplere cu apă la sterilizator (a se vedea Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Când sistemul de umplere cu apă a încercat să umple rezervorul, apa a fost temporar indisponibilă.
SOLUȚIE
Deoarece umplerea rezervorului de apă este încercată doar o singură dată între două cicluri de execuție, acest eveniment inhibă alimentarea cu apă. Opriți sterilizatorul și apoi porniți-l din nou.
Verificați sistemul extern de alimentare cu apă.
Verificați dacă există scurgeri de apă din sterilizator.
CAUZĂ POSIBILĂ
Senzorul de nivel MIN defect în rezervorul de apă curată.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Avertizare privind mentenanța programată.
CAUZĂ POSIBILĂ
Trebuie efectuat programul de mentenanță.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Atenționare privind salvarea pe USB (fișiere HTML și SCL)
CAUZĂ POSIBILĂ
Dispozitivul de stocare USB este deconectat sau plin.
SOLUȚIE
Verificați prezența și starea stick-ului USB. Dacă problema persistă, contactați un partener de service autorizat W&H. -
Faza de sterilizare (PROCESUL) a unui ciclu de sterilizare a fost mai lungă decât se aștepta.
CAUZĂ POSIBILĂ
Temperatura camerei a scăzut sub pragul minim, iar software-ul a efectuat recuperarea cu succes.
SOLUȚIE
Așteptați finalizarea ciclului. Dacă problema apare frecvent, contactați un partener de service autorizat W&H. -
Sterilizatorul intră în "Modul Stand-by" imediat după pornire.
CAUZĂ POSIBILĂ
Unitatea a fost oprită în timp ce se afla în modul standby.
SOLUȚIE
Apăsați butonul Stand-by pentru a ieși. -
La pornirea unui ciclu, ușa camerei se blochează, dar se redeschide imediat. Apare mesajul "Open the door" (Deschideți ușa).
CAUZĂ POSIBILĂ
Garnitura de etanșare a ușii nu este pusă corect.
SOLUȚIE
Asigurați-vă că garnitura de etanșare a ușii este inserată corect pe întreaga circumferință.
CAUZĂ POSIBILĂ
Ușă blocată de obiecte externe sau de instrumentele încărcate.
SOLUȚIE
Îndepărtați orice obiect care se interpune cu ușa camerei. Verificați dacă există eventuale interferențe între ușă și elementele interioare ale camerei și/sau încărcătura. -
Pete, puncte sau modificări de culoare pe instrumente.
CAUZĂ POSIBILĂ
Instrumente clătite cu apă de la robinet și neuscate înainte de sterilizare. SOLUȚIE
Asigurați-vă că instrumentele sunt clătite cu apă distilată sau demineralizată și complet uscate înainte de a fi introduse în sterilizator.
CAUZĂ POSIBILĂ
Utilizarea apei de proastă calitate sau a apei care conține substanțe chimice.
SOLUȚIE
Goliți ambele rezervoare de apă. Folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare)..
CAUZĂ POSIBILĂ
Reziduuri organice sau chimice pe instrumente
SOLUȚIE
Curățați, clătiți și uscați instrumentele înainte de a le introduce în sterilizator (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare).
CAUZĂ POSIBILĂ
Mizerie în cameră, tăvi, suport de tăvi.
SOLUȚIE
Curățați camera și spălați elementele interioare ale camerei.
CAUZĂ POSIBILĂ
Contact între instrumente din diferite materiale.
SOLUȚIE
Asigurați-vă că instrumentele din materiale diferite nu se ating între ele (aluminiu, oțel carbon sau inox etc.); așezați-le pe tăvi sau casete diferite sau puneți-le în pungi (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Depuneri de calcar pe cameră.
SOLUTION
Curățați camera și folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare). -
La sfârșitul ciclului, există apă reziduală pe instrumente.
CAUZĂ POSIBILĂ
Sterilizatorul nu este poziționat corect la nivel. SOLUȚIE
Amplasați sterilizatorul pe o suprafață plană și nivelată.
CAUZĂ POSIBILĂ
Cameră supraîncărcată.
SOLUȚIE
Respectați limitele maxime de greutate a încărcăturii pentru fiecare tip de încărcătură. Folosiți întotdeauna suportul pentru tăvițe și casete.
CAUZĂ POSIBILĂ
Filtrul camerei de sterilizare este înfundat.
SOLUȚIE
Îndepărtați și curățați filtrul.
CAUZĂ POSIBILĂ
Capacul filtrului de cameră poziționat greșit.
SOLUȚIE
Montați corect capacul filtrului de cameră (consultați capitolul "Întreținerea lunară sau la 50 de cicluri - Curățarea filtrului de cameră" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Încărcătură amplasată incorect.
SOLUȚIE
Respectați recomandările enumerate în capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare. -
Sterilizatorul meu are nevoie de vreo calificare/validare periodică?
Validarea cuprinde, în general, o serie de teste definite în standardele recunoscute la nivel internațional. La W&H, aceste teste încep, pentru fiecare echipament, cu testele din fabrică. În plus, documentația corespunzătoare este inclusă împreună cu echipamentul dumneavoastră. Înainte de punerea în funcțiune a dispozitivului la sediul dumneavoastră, tehnicianul efectuează verificări suplimentare pentru a se asigura că acesta funcționează corect. Apoi, în funcție de legile și directivele naționale/regionale, echipamentul trebuie să fie supus unei așa-numite evaluări reînnoite a performanțelor, la intervale regulate. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Atenționare privind înlocuirea consumabilelor.
CAUZĂ POSIBILĂ
Este necesară înlocuirea unui consumabil.
SOLUȚIE
Comandați consumabilul solicitat (garnitura ușii, filtrul HEPA/bacteriologic) și înlocuiți-l. Consultați capitolul "Întreținere" din Instrucțiuni de utilizare. -
Care este nivelul maxim de zgomot al sterilizatorului Lara, Lara XL and Lexa Plus?
Nivelul maxim de zgomot al sterilizatoarelor Lara este 66,9 dB. Nivelul maxim de zgomot al sterilizatoarelor Lara XL/Lexa Plus este de 70 dB.
-
Există scurgeri de apă în partea din față a sterilizatorului.
-
Lisa (24)
-
Există scurgeri de apă în partea din față a sterilizatorului.
CAUZĂ POSIBILĂ
Scurgeri prin garnitura ușii camerei.
SOLUȚIE
Curățați sau înlocuiți garnitura de etanșare a ușii. Curățați suprafața ușii dinspre cameră.
CAUZĂ POSIBILĂ
Scurgere internă.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Sterilizatorul rămâne oprit.
CAUZĂ POSIBILĂ
Întrerupătorul principal sau întrerupătorul de circuit al rețelei este oprit (OFF).
SOLUȚIE
Activați întrerupătorul principal sau întrerupătorul de rețea (ON).
CAUZĂ POSIBILĂ
Nu există tensiune la priză.
SOLUȚIE
Verificați alimentarea electrică a clădirii.
CAUZĂ POSIBILĂ
Cablul de alimentare nu este conectat corespunzător.
SOLUȚIE
Verificați și conectați corect cablul de alimentare. -
Trebuie să efectuez teste periodice pentru sterilizatoarele?
Sterilizatoarele dispun de cicluri de testare (testul B&D/Helix, testul Vacuum). Obligativitatea și frecvența testelor sunt reglementate de directivele locale/naționale.
-
Ar trebui să pun în pungă/împachetez instrumentele și cât timp pot să le păstrez după sterilizare?
Pentru a vă asigura că instrumentele dvs. sunt păstrate în stare sterilă de la sterilizarea lor până la utilizare, W&H recomandă să le împachetați, utilizând sisteme de bariere sterile adecvate (pungi/role de sterilizare) înainte de sterilizare. Perioada de depozitare recomandată poate fi găsită în recomandările producătorului materialului de împachetare. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Ce este validarea?
Validarea este o procedură de testare, efectuată de o autoritate oficială, care dovedește eficiența sterilizatorului dumneavoastră. Obligativitatea și frecvența validării sunt reglementate de normele locale/naționale în vigoare.
-
Ce calitate a apei ar trebui să folosesc pentru sterilizatorul meu?
Calitatea apei pentru sterilizatoare este specificată la nivel internațional prin standarde și directive. Acest lucru se datorează faptului că sterilizatorul dvs. poate fi grav deteriorat de aceasta sau sterilizarea poate fi compromisă. Prin urmare, pentru sterilizatorul dumneavoastră W&H trebuie să utilizați numai apă cu o conductivitate mai mică de 10 μS/cm (în America de Nord) și de 15 μS/cm în restul lumii.Dispozitivele de tratare a apei Multidem și Osmo sunt special concepute pentru a asigura o apă adecvată (a se vedea capitolul "Calitatea apei" din instrucțiunile de utilizare).
-
Cum se curăță rezervoarele de apă și ce produse ar trebui să folosesc?
Instrucțiunile de curățare a rezervorului de apă sunt furnizate în Instrucțiunile de utilizare, capitolul " Mentenanță". În plus, este disponibilă o descriere video: • în sterilizator, în secțiunea de mentenanță din meniu; • în secțiunea Consumabile și accesorii și secțiunea de Tutoriale Video a acestui website.
-
Cum pot schimba componentele consumabile ale sterilizatorului meu?
Instrucțiunile de înlocuire a consumabilelor sunt furnizate în Instrucțiunile de utilizare, capitolul " Mentenanță". În plus, este disponibilă o descriere video: • în sterilizator, în secțiunea de mentenanță din meniu; • în secțiunea Consumabile și accesorii și în secțiunea de Tutoriale Video a acestui website.
-
Pe cine ar trebui să contactez pentru validarea și mentenanța sterilizatorului meu?
Validarea se efectuează de către o autoritate oficială. Sarcinile de mentenanță trebuie efectuate de un tehnician calificat și autorizat.
-
Trebuie să documentez rezultatele tuturor testelor periodice?
W&H recomandă consemnarea tuturor testelor, reparațiilor și modificărilor periodice ale sterilizatorului într-un Jurnal, inclusiv data și semnătura persoanei care a efectuat testul sau altă activitate. Înregistrarea obligatorie este reglementată de normele locale/naționale.
-
Pentru cât timp este necesar să arhivez rapoartele testelor periodice?
Păstrați rezultatele în conformitate cu legile și ghidurile specifice țării/ regiunii dumneavoastră. Aceste registre pot fi utile și în eventualitatea intervențiilor de service.
-
În Meniu, la secțiunea Istoric cicluri, nu există nicio informație despre cicluri.
CAUZĂ POSIBILĂ
S-a înlocuit o placă electronică în timpul reparației.
SOLUȚIE
Memoria vechii plăci nu poate fi recuperată. Consultați istoricul ciclurilor de pe unitatea de memorie USB. -
Imprimanta pentru rapoartele de ciclu nu funcționează.
CAUZĂ POSIBILĂ
Imprimanta nu este conectată corect la sterilizator sau nu este conectată la rețeaua electrică.
SOLUȚIE
Verificați conexiunea de date și de alimentare a imprimantei. -
Sterilizatorul este conectat la un sistem automat de alimentare cu apă, nu există apă curată în rezervor, iar funcția de umplere automată a apei nu alimentează cu apă.
CAUZĂ POSIBILĂ
Sistemul de umplere cu apă nu este conectat.
SOLUȚIE
Conectați sistemul de umplere cu apă la sterilizator (a se vedea Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Când sistemul de umplere cu apă a încercat să umple rezervorul, apa a fost temporar indisponibilă.
SOLUȚIE
Deoarece umplerea rezervorului de apă este încercată doar o singură dată între două cicluri de execuție, acest eveniment inhibă alimentarea cu apă. Opriți sterilizatorul și apoi porniți-l din nou.
Verificați sistemul extern de alimentare cu apă.
Verificați dacă există scurgeri de apă din sterilizator.
CAUZĂ POSIBILĂ
Senzorul de nivel MIN defect în rezervorul de apă curată.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Avertizare privind mentenanța programată.
CAUZĂ POSIBILĂ
Trebuie efectuat programul de mentenanță.
SOLUȚIE
Contactați un partener de service autorizat W&H. -
Atenționare privind salvarea pe USB (fișiere HTML și SCL)
CAUZĂ POSIBILĂ
Dispozitivul de stocare USB este deconectat sau plin.
SOLUȚIE
Verificați prezența și starea stick-ului USB. Dacă problema persistă, contactați un partener de service autorizat W&H. -
Faza de sterilizare (PROCESUL) a unui ciclu de sterilizare a fost mai lungă decât se aștepta.
CAUZĂ POSIBILĂ
Temperatura camerei a scăzut sub pragul minim, iar software-ul a efectuat recuperarea cu succes.
SOLUȚIE
Așteptați finalizarea ciclului. Dacă problema apare frecvent, contactați un partener de service autorizat W&H. -
Sterilizatorul intră în "Modul Stand-by" imediat după pornire.
CAUZĂ POSIBILĂ
Unitatea a fost oprită în timp ce se afla în modul standby.
SOLUȚIE
Apăsați butonul Stand-by pentru a ieși. -
La pornirea unui ciclu, ușa camerei se blochează, dar se redeschide imediat. Apare mesajul "Open the door" (Deschideți ușa).
CAUZĂ POSIBILĂ
Garnitura de etanșare a ușii nu este pusă corect.
SOLUȚIE
Asigurați-vă că garnitura de etanșare a ușii este inserată corect pe întreaga circumferință.
CAUZĂ POSIBILĂ
Ușă blocată de obiecte externe sau de instrumentele încărcate.
SOLUȚIE
Îndepărtați orice obiect care se interpune cu ușa camerei. Verificați dacă există eventuale interferențe între ușă și elementele interioare ale camerei și/sau încărcătura. -
Pete, puncte sau modificări de culoare pe instrumente.
CAUZĂ POSIBILĂ
Instrumente clătite cu apă de la robinet și neuscate înainte de sterilizare. SOLUȚIE
Asigurați-vă că instrumentele sunt clătite cu apă distilată sau demineralizată și complet uscate înainte de a fi introduse în sterilizator.
CAUZĂ POSIBILĂ
Utilizarea apei de proastă calitate sau a apei care conține substanțe chimice.
SOLUȚIE
Goliți ambele rezervoare de apă. Folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare)..
CAUZĂ POSIBILĂ
Reziduuri organice sau chimice pe instrumente
SOLUȚIE
Curățați, clătiți și uscați instrumentele înainte de a le introduce în sterilizator (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare).
CAUZĂ POSIBILĂ
Mizerie în cameră, tăvi, suport de tăvi.
SOLUȚIE
Curățați camera și spălați elementele interioare ale camerei.
CAUZĂ POSIBILĂ
Contact între instrumente din diferite materiale.
SOLUȚIE
Asigurați-vă că instrumentele din materiale diferite nu se ating între ele (aluminiu, oțel carbon sau inox etc.); așezați-le pe tăvi sau casete diferite sau puneți-le în pungi (a se vedea capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Depuneri de calcar pe cameră.
SOLUTION
Curățați camera și folosiți apă de bună calitate (a se vedea capitolul "Calitatea apei" din Instrucțiunile de utilizare). -
La sfârșitul ciclului, există apă reziduală pe instrumente.
CAUZĂ POSIBILĂ
Sterilizatorul nu este poziționat corect la nivel. SOLUȚIE
Amplasați sterilizatorul pe o suprafață plană și nivelată.
CAUZĂ POSIBILĂ
Cameră supraîncărcată.
SOLUȚIE
Respectați limitele maxime de greutate a încărcăturii pentru fiecare tip de încărcătură. Folosiți întotdeauna suportul pentru tăvițe și casete.
CAUZĂ POSIBILĂ
Filtrul camerei de sterilizare este înfundat.
SOLUȚIE
Îndepărtați și curățați filtrul.
CAUZĂ POSIBILĂ
Capacul filtrului de cameră poziționat greșit.
SOLUȚIE
Montați corect capacul filtrului de cameră (consultați capitolul "Întreținerea lunară sau la 50 de cicluri - Curățarea filtrului de cameră" din Instrucțiunile de utilizare).
CAUZĂ POSIBILĂ
Încărcătură amplasată incorect.
SOLUȚIE
Respectați recomandările enumerate în capitolul "Întreținerea și pregătirea încărcăturii" din Instrucțiuni de utilizare. -
Sterilizatorul meu are nevoie de vreo calificare/validare periodică?
Validarea cuprinde, în general, o serie de teste definite în standardele recunoscute la nivel internațional. La W&H, aceste teste încep, pentru fiecare echipament, cu testele din fabrică. În plus, documentația corespunzătoare este inclusă împreună cu echipamentul dumneavoastră. Înainte de punerea în funcțiune a dispozitivului la sediul dumneavoastră, tehnicianul efectuează verificări suplimentare pentru a se asigura că acesta funcționează corect. Apoi, în funcție de legile și directivele naționale/regionale, echipamentul trebuie să fie supus unei așa-numite evaluări reînnoite a performanțelor, la intervale regulate. Pentru informații suplimentare, vă rugăm să contactați dealerul dumneavoastră.
-
Atenționare privind înlocuirea consumabilelor.
CAUZĂ POSIBILĂ
Este necesară înlocuirea unui consumabil.
SOLUȚIE
Comandați consumabilul solicitat (garnitura ușii, filtrul HEPA/bacteriologic) și înlocuiți-l. Consultați capitolul "Întreținere" din Instrucțiuni de utilizare. -
What is the maximum noise level of the Lisa sterilizer?
The max. noise level of Lisa 2019 is 64.3 dB.
-
Există scurgeri de apă în partea din față a sterilizatorului.
Micromotoare Pneumatice
-
Does the air motor need to be greased?
Yes, the air motor must be greased as described in the "Hygiene and care" section of the instructions for use.