Advanced connectivity
The ioDent® system offers intelligent dental solutions and enables an advanced service support.
W&H Steri App
The W&H Steri App is an advanced smartphone and tablet application that allows controlling and remote monitoring of the sterilizer in your dental clinic. Moreover, cycle history back-up automatically saved on smartphone ensures extra-safety.
Compatibility:
The application is compatible with all sterilizers currently available.
Available languages: EN, IT, FR, DE, ES, ZH-CN, JP
Main features:
Remote control of your sterilizer: the user can remotely command up to 4 sterilizers.
Advanced traceability: EliTrace feature is now fully supported and the App permits to create instrument database directly from your smart device. In addition to this function, the management and tracking of dental instruments is integrated.
Reports creation: the App allows making intervention reports in order to collect data regarding specific treatments.
User management: creation and handling of sterilizer users directly from your smart device.
Backup functionality: all the cycle reports are automatically saved to assure more protection of data.Six steps to your customised sterilizer:
- Purchase the Activation Code of your choice from your local approved dental supplier.
- After receiving the related login code, go to activation.wh.com
- Specify this code along with the serial number and model of your sterilizer.
- Your individual, secure Activation Code is generated.
- Enter this code in the menu of your sterilizer.
- The feature is available.
Appropriate accessories
W&H AIMS hygiene workflow
Sterilization, hygiene, and maintenance are increasingly important in the dental practice. The new Lisa Mini is an optimally matching part of W&H´s wide range of products for the safe reprocessing of instruments, contra-angle and turbine handpieces according to the recommended, holistic W&H AIMS hygiene workflow.
W&H AIMS
Product & consumables
Lisa Mini sterilizer, 3 litres
supplied with 2 aluminium trays, drain hose, tray holder, 8 GB USB stick, door opening tool, factory test, compressed air tube,
50/60 HzLisa Mini RIS-303 (220 V)
REF 19944100Chamber size 3 litre, compressed air
Lisa Mini RIS-303 (110 V)
REF 19944200Chamber size 3 litre, compressed air
Lisa Mini sterilizer, 5 litres
supplied with 2 aluminium trays, drain hose, tray holder, 8 GB USB stick, door opening tool, factory test, compressed air tube,
50/60 HzLisa Mini RIS-305 (220 V)
REF 19945100Chamber size 5 litre, compressed air
Lisa Mini RIS-305 (220 V)
REF 19935100Chamber size 5 litre, vacuum pump
Download Area
FAQs
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines. more
-
Who should I use to carry out validation and maintenance on my Lisa/Lina?
All validation and maintenance should be undertaken by the Competent Person (Decontamination) or Competent Person (maintenance). more
-
Does my Lisa/Lina need to be commissioned from new?
Yes, validation is needed for new equipment at installation and annually thereafter. It will also be necessary to validate equipment after any major repairs have been carried out. more
-
How long should I Keep periodic test records?
Under current national decontamination guildance, all audit documents should be stored for at least two years and they should not be removed from the premises or destroyed. more
-
Do I need to document the results of all periodic tests?
Yes, all tests performed by the operator or user should be recorded in a logbook with the date and signature of the individual how carried out the test. Every Sterilizer and every thermo washer disinfector should have a logbook (file) in which test details pertaining to the lifecycle of the equipment (from purchase - disposal) are recorded. A logbook is provided as standard with the Premium Careplus Service Contract from W&H, or as an optional extra with the Premium Care Service Contract. more
-
What is validation?
Validation is a series of tests carried out with the use of independent specialist test equipment, as is necessary to demonstrate that the physical conditions required for sterilization, thermal disinfection and/or cleaning (i.e. temperature, pressure, time) are achieved. more
-
What water quality should I use with my Lisa/Lina?
The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardised. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Water treatment devices Multidem and Osmo are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use). more
-
Should I pre-wrap and how long can I store my instruments after sterilization?
Under current UK decontamination guidance, with a type B vacuum cycle, instruments should be pre-wrapped and once sterilized can be stored for up to 12 months. Wrapping should take place shortly after washing and disinfection. Under current UK decontamination guidance, with a type N cycle, instruments should not be pre-wrapped. Once sterilized, instruments can be immediately aseptically wrapped and stored for up to 12 months. Instruments should always be dry before they are placed in the purpose-designed packaging. more
-
What is the minimum shaft length of the bur for the S-11 Surgical Handpiece?
The shaft of the bur should be at least 44mm long (ISO-shaft 104). more
-
What is the maximum noise level of the Lisa Mini sterilizer?
The max. noise level of Lisa Mini is 64 dB. more
-
How do I change consumable components on my sterilizer?
Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu. more
-
How do I clean the water tank and what products should I use?
Water tank cleaning instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu. more
-
Is Lisa Mini requiring compressed air connection?
Lisa Mini requires compressed air connection to perform sterilization cycle. The compressed air specifications are available in the Instruction for Use, chapter “Technical data”. more
Instructions for use
Find all languages of the Update Hygiene and Maintenance (EN ISO 17664:2017) in the Download Centre.
Video Tutorials
-
Lisa Mini sterilization cycles
Sterilization cycle: B Universal 134 B Prion 134 B Universal 121 S Wrapped 134 S Unwrapped Temperature: 134°C 134°C 121°C 134°C 134°C Holding time (minutes): 4' 18' 30" 20' 30" 4' 3' 30" Total cycle duration
(minutes)
including drying:3l (1) 19' 45"–23' 45" 34' 15"–38' 15" 37' 15"–38' 45" 17' 45"–19' 30" 8' 45"–10' 30" 5l (1) 21' 45"–29' 15" 36' 15"–43' 45" 38' 15"–45' 15" 18' 45"–22' 45" 9' 45"–13' 45" 5l (2) 23' 45"–31' 15" 38' 15"–45' 45" 39' 45"–46' 45" 20' 45"–24' 45" 10' 15"–14' 15" Load type: All types of load (3) Solid items, Simple hollow items (Hollow B), Specific load: dental load, medical load (4) and handpieces Max. load: Wrapped and unwrapped items: 1 kg (3 litre)/1.5 kg (5 litre) Unwrapped items:
1 kg (3 litre)/
1.5 kg (5 litre)Wrapped and unwrapped porous/textile: 100 g unwrapped items: 1 kg
(3 litre)/1.5 kg (5 litre)Test cycles: Helix/Bowie and Dick/Vacuum Item supplied with 2 aluminium trays, drain hose, tray holder, 8 GB USB stick, door opening tool, factory test, compressed air tube (1) . (1) Compressed air, (2) Vacuum pump, (3) Excluded liquid
(4) Dental load according to ISO 17665-3; for medical load is intended to load up to grade 7 steam penetration according to ISO 17665-3.Lisa Mini
Trade name:
Type:Lisa Mini
RIS-303Lisa Mini
RIS-305Chamber size: 3 litres 5 litres Power supply: 200–240 V AC; 50/60 Hz; 10 A single phase
100–125 V AC; 50/60 Hz; 12 A single phase200–240 V AC; 50/60 Hz; 10 A single phase Power consumption: 2–2.4 kW (220 V)/1.2–1.5 kW (110 V) Overall dimension (w × h × d): 440 mm × 200 mm × 581 mm 440 mm × 200 mm × 666 mm Weight: 21.9 kg 24.4 kg Main/used water tanks: 1.6 l/0.6 l 2.5 l/0.6 l Working range: 3 to 7 cycles Average air consumption: 50–60 (N)l/min (13.2–15.8 (N)gal/min) Working pressure: 5–10 bar (73–145 psi) Connection types: 5 USB ports, 1 Ethernet port, automatic water filling kit Lisa Mini sterilizers were designed, certified and validated with the most stringent directives and standards: 2017/745
Medical Device RegulationIEC 61326-1
Electromagnetic compatibilityIEC 61770
Electric appliances connected to the water mains2019/19/EU
Waste Electrical and Electronic EquipmentIEC 61010-1
Safety requirementsEN 13060
Small steam sterilizers2014/68/EU
Pressure Equipment DirectiveIEC 61010-2-040
Specific requirements for steam sterilizer> The sterilizer can be validated according
to EN ISO 17665-1