Reliable and safe sterilization with Lyla
In the reprocessing workflow, the sterilization of your dental instruments is deeply important.
That is why we offer the highest standards.Lyla reduces the risk
of infection through its- Smart upgrades
- Super fast cycles
- Secure traceability
-
Simple operation
-
New environmentally friendly packaging
The new environmentally friendly packaging is completely reusable and recyclable. Plastic components were reduced to a minimum. For 99% of its weight, the new package is made of cardboard and other fully biodegradable substances.
The "B cycle”
feature upgrades Lyla to B type sterilizer according to
EN 13060.Upgradeable from S to B
The “B cycle” feature upgrades Lyla simply
and quick to a B type sterilizer according to
EN 13060.
Remote data storage
offers the possibility to save cycles directly on PC.Remote data Storage
The “Remote data storage” feature
is the ideal addition for the digital
cycle documentation.
The “Traceability”
feature enables user identification and load release confirmation.Traceability
Identification is secured by a PIN
directly on the Lyla sterilizer and
the W&H Steri App.Six steps to your customized sterilizer:
- Purchase Activation Code of your choice from your local approved dental supplier.
- After receiving the related login code go to activation.wh.com.
- Specify this code along with the serial number and model of your sterilizer.
- Your individual, secure Activation Code is generated.
- Enter this code in the menu of your Lyla sterilizer.
- The feature is available.
Compatibility:
The application is compatible with all sterilizers currently available.
Available languages: EN, IT, FR, DE, ES, ZH-CN, JP
Main features:
Remote control of your sterilizer: the user can remotely command up to 4 sterilizers.
Advanced traceability: EliTrace feature is now fully supported and the App permits to create instrument database directly from your smart device. In addition to this function, the management and tracking of dental instruments is integrated.
Reports creation: the App allows making intervention reports in order to collect data regarding specific treatments.
User management: creation and handling of sterilizer users directly from your smart device.
Backup functionality: all the cycle reports are automatically saved to assure more protection of data.
Lyla sterilizer
supplied with reversible rack, 3 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test,
50 HzLyla 17
REF 19951150Chamber size 17 litre
Lyla 22 (220 V)
REF 19952150Chamber size 22 litre
Lyla Sterilizer
supplied with reversible rack, 3 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test,
60 HzLyla 17
REF 19951250Chamber size 17 litre
Lyla 22 (220 V)
REF 19952250Chamber size 22 litre
Download Area
FAQs
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines. more
-
Who should I use to carry out validation and maintenance on my Lisa/Lina?
All validation and maintenance should be undertaken by the Competent Person (Decontamination) or Competent Person (maintenance). more
-
Does my Lisa/Lina need to be commissioned from new?
Yes, validation is needed for new equipment at installation and annually thereafter. It will also be necessary to validate equipment after any major repairs have been carried out. more
-
How long should I Keep periodic test records?
Under current national decontamination guildance, all audit documents should be stored for at least two years and they should not be removed from the premises or destroyed. more
-
Do I need to document the results of all periodic tests?
Yes, all tests performed by the operator or user should be recorded in a logbook with the date and signature of the individual how carried out the test. Every Sterilizer and every thermo washer disinfector should have a logbook (file) in which test details pertaining to the lifecycle of the equipment (from purchase - disposal) are recorded. A logbook is provided as standard with the Premium Careplus Service Contract from W&H, or as an optional extra with the Premium Care Service Contract. more
-
What is validation?
Validation is a series of tests carried out with the use of independent specialist test equipment, as is necessary to demonstrate that the physical conditions required for sterilization, thermal disinfection and/or cleaning (i.e. temperature, pressure, time) are achieved. more
-
What water quality should I use with my Lisa/Lina?
The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardised. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Water treatment devices Multidem and Osmo are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use). more
-
How do I clean the water tanks and what products should I use?
Water tank cleaning instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu; • in the Consumables & Accessories and Video Tutorial section of this website. more
-
How do I change consumable components on my sterilizer?
Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu; • in the Consumables & Accessories and Video Tutorial section of this website. more
-
Should I pre-wrap and how long can I store my instruments after sterilization?
Under current UK decontamination guidance, with a type B vacuum cycle, instruments should be pre-wrapped and once sterilized can be stored for up to 12 months. Wrapping should take place shortly after washing and disinfection. Under current UK decontamination guidance, with a type N cycle, instruments should not be pre-wrapped. Once sterilized, instruments can be immediately aseptically wrapped and stored for up to 12 months. Instruments should always be dry before they are placed in the purpose-designed packaging. more
-
Lisa 17/22<br />
At the end of the cycle, there is residual water in the chamber
POSSIBLE CAUSE
Sterilizer not properly levelled
SOLUTION
Properly level the surface the sterilizer is placed on.
POSSIBLE CAUSE
Overloaded chamber
SOLUTION
Comply with the maximum load weight limits for each type of load. Always use the chamber rack for trays and cassettes.
POSSIBLE CAUSE
Chamber filter clogged
SOLUTION
Remove and clean the chamber filter.
POSSIBLE CAUSE
Chamber filter cap mispositioned
SOLUTION
Mount the chamber filter cap properly (see chapter MAINTENANCE)
POSSIBLE CAUSE
Load incorrectly placed
SOLUTION
Follow the recommendations as listed in ANNEX 2. more -
Lisa 17/22<br />
Water is leaking at the front of the sterilizer
POSSIBLE CAUSE
Leaks through the chamber door seal
SOLUTION
Clean or replace the door seal. Clean the chamber face side.
POSSIBLE CAUSE
Internal leak.
SOLUTIONS
Call technical service. more -
Lisa 17/22<br />
The sterilizer remains switched OFF.
POSSIBLE CAUSE
The main switch or network circuit breaker is OFF
SOLUTION
Activate the main switch or network circuit breaker (ON).
POSSIBLE CAUSE
No voltage at the socket
SOLUTION
Check the electric circuit.
POSSIBLE CAUSE
The power cord is not connected properly
SOLUTION
Check and connect the power cord properly. more -
Lisa 17/22<br />
The sterilizer enters into “Sleep mode” immediately after opening the chamber door.
POSSIBLE CAUSE
The chamber door has not been opened after the previous cycle had finished and the “Sleep mode delay” has expired
SOLUTION
Press the SLEEP MODE button to exit”. more -
Lisa 17/22<br />
When starting a cycle, the chamber door locks but re-opens immediately. The “Open the door” message appears.
POSSIBLE CAUSE
Door seal not properly placed; seal sticking out
SOLUTION
Ensure that the door seal is correctly inserted around the whole circumference.
POSSIBLE CAUSE
Door jammed by external objects or by the load itself
SOLUTION
Remove any objects interfering with the chamber door. Check that the door does not push against the load or the chamber furniture. more -
Lisa 17/22<br />
Corrosion or spots on instruments
POSSIBLE CAUSE
Tap water on instruments when placed in the sterilizer
SOLUTION
Ensure that instruments are dry before they are placed in the sterilizer.
POSSIBLE CAUSE
Use of water of poor quality or water containing chemical substances
SOLUTION
Drain both water tanks. Use water of good quality (see ANNEX 7).
POSSIBLE CAUSE
Organic or chemical residues on the instruments
SOLUTION
Clean, rinse and dry instruments before placing them in the sterilizer (see ANNEX 2).
POSSIBLE CAUSE
Chamber, trays, tray rack dirty
SOLUTION
Clean the chamber and wash the chamber furniture
POSSIBLE CAUSE
Contact between instruments of different materials
SOLUTION
Ensure that instruments of different materials do not touch (aluminum, carbon or stainless steel, etc.); place them on different trays or cassettes or pouch them (refer to ANNEX 2).
POSSIBLE CAUSE
Scale deposits on the chamber
SOLUTION
Clean the chamber and use water of good quality (refer to ANNEX 7). more -
Lisa 17/22<br />
No cycles are stored in the cycle history menu
POSSIBLE CAUSE
An electronic board was replaced by service.
SOLUTION
None. The memory of the old board cannot be restored. Save periodically the history on the USB pen drive. more -
Lisa 17/22<br />
The sterilization (PROCESS) phase of a sterilization cycle was longer than expected.
POSSIBLE CAUSE
The chamber temperature dropped below the minimum threshold and the software performed a successful recovery.
SOLUTION
Wait for cycle completion. If the problem occurs frequently, call technical service. more -
Warning about saving to the USB (HTML and SCL files).
The USB drive is disconnected or full. SOLUTION Check presence and condition of the USB drive. If the problem persists, contact an authorised W&H service partner. more
-
Lisa 17/22<br />
The cycle report printer does not work.
POSSIBLE CAUSE
Printer not properly connected or not powered
SOLUTION
Check the data and the power connection to the printer. more -
Lisa 17/22<br />
Warning about programmed maintenance.
POSSIBLE CAUSE
A component shall be replaced for the programmed maintenance of the sterilizer.
SOLUTION
Call service to order the requested component (door seal, dust filter, bacteriological filter...). See “Maintenance” chapter.. more -
Lisa 17/22<br />
When the sterilizer is connected to an automated water supply system: there is no clean water in the tank, but the automatic water filling does not fill the water.
POSSIBLE CAUSE
Water fill system not connected
SOLUTION
Connect the water fill system to the sterilizer (see ANNEX 7 for water quality requirements).
POSSIBLE CAUSE
When the water fill system attempted to fill the tank, water was temporarily unavailable
SOLUTION
Since water tank filling is attempted only once in-between cycle execution, this event inhibits water feeding. Switch the sterilizer OFF and then ON again.
Check the external water supply system.
Check for water leaks from the sterilizer.
POSSIBLE CAUSE
Faulty MIN water level sensor in the clean water tank
SOLUTION
Call service. more -
What should I do if I get a Warning about consumable replacement?
Order the requested consumable (door seal and HEPA/bacteriological filter) and replace it. See the chapter “Maintenance” in the Instructions for Use. more
-
Which is the maximum noise level of Lyla?
The max. noise level of Lyla is 65.5 dB. more
Instructions for use
Find all languages of the Update Hygiene and Maintenance (EN ISO 17664:2017) in the Download Centre.
Video Tutorials
-
Lyla sterilization programs
Sterilization cycle S Wrapped 134 S Prion 134 S Gentle 121 S Unwrapped 134 Temperature: 134 °C 134 °C 121 °C 134 °C Holding time (minutes): 4 18' 30'' 20' 30'' 4 Total cycle duration
(minutes) including
complete drying:Automatic from Typical
loadAutomatic from Typical
loadAutomatic from Automatic from Typical
loadLyla 17 38' to 45'42' 52' to 59'56' 61' to 64' 20' to 26' 23' Lyla 22 (220 V) 39' to 46'43' 53' to 60'57' 62' to 66' 21' to 27' 24' Lyla 22 (110 V) 49' to 56'51' 63' to 71'65' 69' to 73
28' to 35' 30' Lyla 17/22 loads:maximum solid 4 kg/typical 2 kg/maximum porous 0.5 kgmaximum 4 kg unwrappedLyla 17/22 container load: maximum 4 kg not applicable Test cycles: Bowie and Dick/Vacuum test Item supplied with reversible rack, 3 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test
Lyla Sterilizer
Lyla 17 Lyla 22 Type: RIN-210 RIN-210 Chamber size: 17 l 22 l Power supply: 200–240 V AC; 50 Hz or 60 Hz; 10 A single phase
200–240 V AC; 50 Hz or 60 Hz; 10 A single phase
100–125 V AC; 50 Hz/60 Hz; 12 A single phase
Power consumption: 2.0–2.4 kW 220 V: 2.0–2.4 kW
110 V: 1.2–1.5 kWOverall dimension
(width × height × depth):465 × 452 × 646 mm Weight: 40 kg 42 kg Main / used water tanks: 4.8/4.8 l Working range: 7 to 10 cycles Usable space in chamber:
(width × height × depth)¹:195 × 195 × 312 mm 195 × 195 × 400 mm Connection types: 1 USB port in the front (additional 1 in the rear as option), automatic water filling kit (optional) (1) Usable space is referred to standard rack configuration.
Lisa sterilizers were designed, certified and validated with the most stringent directives and standards: N°2017/745 Medical Device Regulation 2014/68/EU Pressure Equipment Directive EN 13060 Small steam sterilizer IEC 61326-1 Electromagnetic compatibility IEC 61010-1 Safety requirements sterilizer
IEC 61010-2-040 Specific requirements for steam sterilizer
IEC 61770 Electric appliances connected to the water mains
The sterilizer can be validated according to EN ISO 17665-1